Analytical verification of 12 most commonly used urine dipsticks in Croatia: comparability, repeatability and accuracy
Analytical verification of 12 most commonly used urine dipsticks in Croatia: comparability, repeatability and accuracy // Biochemia Medica, 29 (2019), 1; 010708, 10 doi:10.11613/BM.2019.010708 (domaća recenzija, članak, znanstveni)
CROSBI ID: 984922 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Analytical verification of 12 most commonly used urine dipsticks in Croatia: comparability, repeatability and accuracy
Autori
Vuljanić, Dora ; Dojder, Ana ; Špoljarić, Valentina ; Saračević, Andrea ; Dukić, Lora ; Leniček-Krleža, Jasna ; Vlašić-Tanasković, Jelena ; Maradin, Ivana ; Grzunov, Ana ; Vogrinc, Željka ; Šimundić, Ana-Maria
Izvornik
Biochemia Medica (1330-0962) 29
(2019), 1;
010708, 10
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
verification ; urine dipsticks ; comparability ; accuracy ; repeatability
Sažetak
Variability among manufacturers of urine dipsticks, respective to their accuracy and measurement range, may lead to diagnostic errors and thus create a serious risk for the patient. Our aims were to determine the level of agreement between 12 most commonly used urine dipsticks in Croatia, examine their accuracy for glucose and total protein and to test their repeatability. A total of 75 urine samples were used to examine comparability and accuracy of 12 dipstick brands (Combur 10 TestM, ChoiceLine 10, Combur 10 TestUX, ComboStik 10M, ComboStik 11M, CombiScreen 11SYS, CombiScreen 10SL, Combina 13, Combina 11S, Combina 10M, UriGnost 11, Multistix 10SG). Agreement between each dipstick and the reference (Combur 10 TestM) was expressed as kappa coefficient (acceptable κ ≥ 0.80). Accuracy for glucose and total protein was tested by comparison with quantitative measurements on analysers: AU400 (Beckman Coulter, USA), Cobas 6000 c501 (Roche Diagnostics, Germany) and Architect plus c4000 (Abbott, USA). Repeatability was assessed on 20 replicates (acceptable > 90%). Best agreement was achieved for glucose, total protein and nitrite (11/11, k > 0.80) and the lowest for bilirubin (5/5, k < 0.60). Sensitivities for total protein were 41- 75% (AU400) and 56-92% (Cobas and Architect) ; while specificities were 41-75% (AU400, Cobas, Architect). Dipsticks’ sensitivity and specificity for glucose were 68-98%. Most of the dipsticks showed unacceptable repeatability (6/12, < 90%) for one parameter, most prominently for pH (3/12, < 90%). Most commonly used dipsticks in Croatia showed low level of agreement between each other. Moreover, their repeatability varies among manufacturers and their accuracy for glucose and proteins is poor.
Izvorni jezik
Engleski
Znanstvena područja
Kliničke medicinske znanosti
POVEZANOST RADA
Ustanove:
Klinička bolnica "Sveti Duh"
Profili:
Željka Vogrinc
(autor)
Ana-Maria Šimundić
(autor)
Andrea Saračević
(autor)
Lora Dukić
(autor)
Jasna Leniček Krleža
(autor)
Citiraj ovu publikaciju:
Časopis indeksira:
- Current Contents Connect (CCC)
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus
- MEDLINE
Uključenost u ostale bibliografske baze podataka::
- EMBASE (Excerpta Medica)
- EBSCO
- JCR
- ScienceCentral