Pregled bibliografske jedinice broj: 98369
Analysis of Coupling between 900 MHz Mobile Telephone and Implanted Pacemaker
Analysis of Coupling between 900 MHz Mobile Telephone and Implanted Pacemaker // Proceeding on European Microwave Conference 2001 / Nightingale, S. ; Philippakis, M. (ur.).
London : Delhi: European Microwave Association, 2001. str. 365-368 (predavanje, međunarodna recenzija, cjeloviti rad (in extenso), znanstveni)
CROSBI ID: 98369 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Analysis of Coupling between 900 MHz Mobile Telephone and Implanted Pacemaker
Autori
Hrabar, Silvio ; Balachandran, Wamadeva ; Bartolić, Juraj
Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni
Izvornik
Proceeding on European Microwave Conference 2001
/ Nightingale, S. ; Philippakis, M. - London : Delhi : European Microwave Association, 2001, 365-368
Skup
European Microwave Conference
Mjesto i datum
London, Ujedinjeno Kraljevstvo, 25.09.2001. - 27.09.2001
Vrsta sudjelovanja
Predavanje
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
pacemaker; electromagnetic compatibility; interference; mobile telephone
Sažetak
The worst-case electromagnetic coupling between a 900 MHz mobile telephone and implanted pacemaker with unipolar catheter was investigated by computer simulation. The model comprises a dielectric block with embedded pacemaker and a half-wavelength dipole situated above the surface. A pacemaker connector was modelled as a small loop and a catheter as an insulated linear antenna. This model was combined with Finite Difference Time Domain calculation of the field induced within dielectric block. The analysis with 2W transmitter gave the peak interference voltage of 1 V for the connector and 0.3 V for the catheter. These results show that the pacemaker connector is more responsible for coupling than a catheter.
Izvorni jezik
Engleski
Znanstvena područja
Elektrotehnika