Pregled bibliografske jedinice broj: 883218
Titration of anionic surfactants using new potentiometric sensors
Titration of anionic surfactants using new potentiometric sensors // Analytical letters, 22 (1989), 13/14; 2791-2802 doi:10.1080/00032718908052395 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 883218 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Titration of anionic surfactants using new potentiometric sensors
Autori
Zelenika, Ljerka ; Sak-Bosnar, Milan ; Marek, N. ; Kovacs, Barna
Izvornik
Analytical letters (0003-2719) 22
(1989), 13/14;
2791-2802
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
ion selective electrode, anionic surfactant, potentiometric titration
Sažetak
An anionic surfactant ion selective electrode has been developed by coating the Rutitka Selectrode with plasticised polyvinyl chloride containing hyamine-dodecylsulfate active substance. The electrode responds to dodecylsulfate, in the range 10-2 – 10-6 mol dm-3 and has been used for potentiometric titration of micromolar amounts of dodecylsulfate with an accuracy of 0.3%. The influence of several organic and inorganic anions has been investigated.
Izvorni jezik
Engleski
Znanstvena područja
Kemija
POVEZANOST RADA
Ustanove:
Sveučilište J. J. Strossmayera u Osijeku
Profili:
Milan Sak-Bosnar
(autor)
Citiraj ovu publikaciju:
Časopis indeksira:
- Current Contents Connect (CCC)
- Web of Science Core Collection (WoSCC)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus