Analytical evaluation of the clinical chemistry analyzer Olympus AU2700 plus
Analytical evaluation of the clinical chemistry analyzer Olympus AU2700 plus // Biochemia medica, 20 (2010), 3; 334-340 doi:10.11613/BM.2010.043 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 818853 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Analytical evaluation of the clinical chemistry analyzer Olympus AU2700 plus
Autori
Juriček, Jasna ; Đerek, Lovorka ; Unić, Adriana ; Serdar, Tihana ; Marijančević, Domagoj ; Živković, Marcela ; Romić, Željko
Izvornik
Biochemia medica (1330-0962) 20
(2010), 3;
334-340
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
Olympus AU2700 plus ; analytical evaluation ; imprecision ; inaccuracy ; Passing Bablok regression analysis
Sažetak
The objective of this study was to perform the analytical evaluation of the clinical chemistry analyzer Olympus AU2700 plus. The evaluation was performed according to the guidelines of the European Committee for Clinical Laboratory Standards (ECCLS). The evaluation consisted of determination of within-run and between-run imprecision, inaccuracy and comparison with Olympus AU2700. The tested analytes were: glucose, creatinine, urate, total bilirubin, cholesterol, tryglicerides, calcium, phosphate, iron, unsaturated iron binding capacity, aspartate aminotransferase, alanine aminotransferase, lactate dehydrogenase, creatine kinase, alpha-amylase, alkaline phosphatase, potassium, sodium, chloride, C- reactive protein, antistreptolysin O and rheumatoid factor. The results showed low within-run and between-run imprecision and acceptable inaccuracy for all analytes. High correlation with AU2700 analyzer was found. Intercept and slope (with 95% confidence interval) met the preferences of Passing-Bablok regression for all of the analytes except alanine aminotransferase, sodium, rheumatoid factor, creatine kinase, total bilirubin and tryglicerides. Olympus AU2700 plus analyzer shows acceptable precision and accuracy for majority of analytes, with the exception of sodium and chloride. Instrument is fully comparable with Olympus AU2700 analyzer for all but several analytes (ALT, RF, sodium, CK, total bilirubin and tryglicerides), where only some minor deviations (constant and proportional difference) were observed.
Izvorni jezik
Engleski
Znanstvena područja
Kliničke medicinske znanosti
POVEZANOST RADA
Ustanove:
Klinička bolnica "Dubrava"
Profili:
Lovorka Đerek
(autor)
Jasna Juriček
(autor)
Tihana Serdar Hiršl
(autor)
Adriana Unić
(autor)
Domagoj Marijančević
(autor)
Željko Romić
(autor)
Marcela Živković
(autor)
Citiraj ovu publikaciju:
Časopis indeksira:
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus