CALIBRATION OF OPTICAL MEASURING MACHINE USING LASER INTERFEROMETER


Bašić, Hazim; Softić, Almira; Bošnjaković, Alen
CALIBRATION OF OPTICAL MEASURING MACHINE USING LASER INTERFEROMETER // 15th International Research/Expert Conference ” Trends in the Development of Machinery and Associated Technology” / Dr. Sabahudin Ekinović, Dr. Joan Vivancos Calvet, Dr. Emin Tacer (ur.).
Prag, Češka Republika, 2011. str. 297-300 (predavanje, međunarodna recenzija, cjeloviti rad (in extenso), znanstveni)


CROSBI ID: 805675 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
CALIBRATION OF OPTICAL MEASURING MACHINE USING LASER INTERFEROMETER

Autori
Bašić, Hazim ; Softić, Almira ; Bošnjaković, Alen

Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni

Izvornik
15th International Research/Expert Conference ” Trends in the Development of Machinery and Associated Technology” / Dr. Sabahudin Ekinović, Dr. Joan Vivancos Calvet, Dr. Emin Tacer - , 2011, 297-300

Skup
15th International Research/Expert Conference ” Trends in the Development of Machinery and Associated Technology”

Mjesto i datum
Prag, Češka Republika, 12.09.2011. - 18.09.2011

Vrsta sudjelovanja
Predavanje

Vrsta recenzije
Međunarodna recenzija

Ključne riječi
length standard ; measure uncertainties ; beat frequency ; iodine stabilized He-Ne laser

Sažetak
NPL iodine stabilized He-Ne laser at wavelength 633 nm is intended for work with the 2nd class lasers in visible spectrum, respectively, it is laser with output energy that is lower than 1mW and wavelength from 400 nm to 700 nm. This referent standard is design to be stabilized to the b21 component of the 6-3 P (33) iodine transition. This transition is listed in the CIPM recommendation for practical realization of the meter. NPL iodine stabilized He-Ne laser was used for calibration of He-Ne laser. The technique for dissemination unit of the length on the highest metrology level which is used for calibration of stabilized He-Ne lasers is called beat frequency. In the experimental part of work the practical mode of conducting calibration of frequency stabilized He-Ne Laser, respectively, determination its referent frequency with the estimation uncertainty of measurement was done. Additionally, parameters that have influence on measuring results were estimated.

Izvorni jezik
Engleski

Znanstvena područja
Strojarstvo



POVEZANOST RADA


Ustanove:
Fakultet strojarstva i brodogradnje, Zagreb

Profili:

Avatar Url Alen Bošnjaković (autor)

Citiraj ovu publikaciju:

Bašić, Hazim; Softić, Almira; Bošnjaković, Alen
CALIBRATION OF OPTICAL MEASURING MACHINE USING LASER INTERFEROMETER // 15th International Research/Expert Conference ” Trends in the Development of Machinery and Associated Technology” / Dr. Sabahudin Ekinović, Dr. Joan Vivancos Calvet, Dr. Emin Tacer (ur.).
Prag, Češka Republika, 2011. str. 297-300 (predavanje, međunarodna recenzija, cjeloviti rad (in extenso), znanstveni)
Bašić, H., Softić, A. & Bošnjaković, A. (2011) CALIBRATION OF OPTICAL MEASURING MACHINE USING LASER INTERFEROMETER. U: Dr. Sabahudin Ekinović, Dr. Joan Vivancos Calvet, Dr. Emin Tacer (ur.)15th International Research/Expert Conference ” Trends in the Development of Machinery and Associated Technology”.
@article{article, author = {Ba\v{s}i\'{c}, Hazim and Softi\'{c}, Almira and Bo\v{s}njakovi\'{c}, Alen}, year = {2011}, pages = {297-300}, keywords = {length standard, measure uncertainties, beat frequency, iodine stabilized He-Ne laser}, title = {CALIBRATION OF OPTICAL MEASURING MACHINE USING LASER INTERFEROMETER}, keyword = {length standard, measure uncertainties, beat frequency, iodine stabilized He-Ne laser}, publisherplace = {Prag, \v{C}e\v{s}ka Republika} }
@article{article, author = {Ba\v{s}i\'{c}, Hazim and Softi\'{c}, Almira and Bo\v{s}njakovi\'{c}, Alen}, year = {2011}, pages = {297-300}, keywords = {length standard, measure uncertainties, beat frequency, iodine stabilized He-Ne laser}, title = {CALIBRATION OF OPTICAL MEASURING MACHINE USING LASER INTERFEROMETER}, keyword = {length standard, measure uncertainties, beat frequency, iodine stabilized He-Ne laser}, publisherplace = {Prag, \v{C}e\v{s}ka Republika} }