Pregled bibliografske jedinice broj: 805675
CALIBRATION OF OPTICAL MEASURING MACHINE USING LASER INTERFEROMETER
CALIBRATION OF OPTICAL MEASURING MACHINE USING LASER INTERFEROMETER // 15th International Research/Expert Conference ” Trends in the Development of Machinery and Associated Technology” / Dr. Sabahudin Ekinović, Dr. Joan Vivancos Calvet, Dr. Emin Tacer (ur.).
Prag, Češka Republika, 2011. str. 297-300 (predavanje, međunarodna recenzija, cjeloviti rad (in extenso), znanstveni)
CROSBI ID: 805675 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
CALIBRATION OF OPTICAL MEASURING MACHINE USING LASER INTERFEROMETER
Autori
Bašić, Hazim ; Softić, Almira ; Bošnjaković, Alen
Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni
Izvornik
15th International Research/Expert Conference ” Trends in the Development of Machinery and Associated Technology”
/ Dr. Sabahudin Ekinović, Dr. Joan Vivancos Calvet, Dr. Emin Tacer - , 2011, 297-300
Skup
15th International Research/Expert Conference ” Trends in the Development of Machinery and Associated Technology”
Mjesto i datum
Prag, Češka Republika, 12.09.2011. - 18.09.2011
Vrsta sudjelovanja
Predavanje
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
length standard ; measure uncertainties ; beat frequency ; iodine stabilized He-Ne laser
Sažetak
NPL iodine stabilized He-Ne laser at wavelength 633 nm is intended for work with the 2nd class lasers in visible spectrum, respectively, it is laser with output energy that is lower than 1mW and wavelength from 400 nm to 700 nm. This referent standard is design to be stabilized to the b21 component of the 6-3 P (33) iodine transition. This transition is listed in the CIPM recommendation for practical realization of the meter. NPL iodine stabilized He-Ne laser was used for calibration of He-Ne laser. The technique for dissemination unit of the length on the highest metrology level which is used for calibration of stabilized He-Ne lasers is called beat frequency. In the experimental part of work the practical mode of conducting calibration of frequency stabilized He-Ne Laser, respectively, determination its referent frequency with the estimation uncertainty of measurement was done. Additionally, parameters that have influence on measuring results were estimated.