Pregled bibliografske jedinice broj: 772699
A gas ionisation detector in the axial (Bragg) geometry used for the Time-of-Flight Elastic Recoil Detection Analysis
A gas ionisation detector in the axial (Bragg) geometry used for the Time-of-Flight Elastic Recoil Detection Analysis // Review of scientific instruments, 86 (2015), 083301-1 doi:10.1063/1.4927605 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 772699 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
A gas ionisation detector in the axial (Bragg) geometry used for the Time-of-Flight Elastic Recoil Detection Analysis
Autori
Siketić, Zdravko ; Skukan, Natko ; Bogdanović Radović, Ivančica
Izvornik
Review of scientific instruments (0034-6748) 86
(2015);
083301-1
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
Gas detector; mass resolution; Energy resolution; TOF-ERDA
Sažetak
In this paper Time-of-Flight Elastic Recoil Detection Analysis (TOF-ERDA) spectrometer with a newly constructed gas ionization detector (GID) for energy detection is presented. The detector is designed in the axial (Bragg) geometry with a 3x3 array of 50 nm thick Si3N4 membranes as an entrance window. 40 mbar isobutane gas was sufficient to stop a 30 MeV primary iodine beam as well as all recoils in the detector volume. Spectrometer and detector performances were determined showing significant improvement in the mass and energy resolution respectively, comparing to the spectrometer with a standard silicon particle detector for an energy measurements.
Izvorni jezik
Engleski
Znanstvena područja
Fizika
POVEZANOST RADA
Ustanove:
Institut "Ruđer Bošković", Zagreb
Citiraj ovu publikaciju:
Časopis indeksira:
- Current Contents Connect (CCC)
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus