Pretražite po imenu i prezimenu autora, mentora, urednika, prevoditelja

Napredna pretraga

Pregled bibliografske jedinice broj: 772674

A floating multi-channel picoammeter for micropattern gaseous detector current monitoring


Utrobičić, Antonija; Kovačić, Marinko; Erhardt, Filip; Poljak, Nikola; Planinić, Mirko
A floating multi-channel picoammeter for micropattern gaseous detector current monitoring // Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 801 (2015), 21-26 doi:10.1016/j.nima.2015.08.021 (međunarodna recenzija, članak, znanstveni)


CROSBI ID: 772674 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
A floating multi-channel picoammeter for micropattern gaseous detector current monitoring

Autori
Utrobičić, Antonija ; Kovačić, Marinko ; Erhardt, Filip ; Poljak, Nikola ; Planinić, Mirko

Izvornik
Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment (0168-9002) 801 (2015); 21-26

Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni

Ključne riječi
Floating current measurement ; picoammeter ; digital readout ; GEM detector

Sažetak
This paper presents a concept of a multiple channel picoammeter suitable for operation at a high voltage. To ensure floating operation, 1 kHz readout rate and low battery usage, the picoammeter employs an optical fibre for data transmission. The analogue conditioning section consists of a transimpedance amplifier that is connected to a 16 bit Sigma Delta ADC which is integrated into the microcontroller. Simultaneous reception of data sent from multiple picoammeter units is achieved using an FPGA device and a USB interface to the host PC. A prototype with a +/- 125 nA range was constructed. The paper includes a calibration procedure of the prototype and an example of a current leakage test conducted on a 10x10 cm2 GEM foil. The resolution achieved by the prototype in a 15 min test at 0, 5 nA bias was 6, 5 pA FWHM.

Izvorni jezik
Engleski

Znanstvena područja
Fizika, Elektrotehnika



POVEZANOST RADA


Projekti:
UKF 14/13

Ustanove:
Fakultet elektrotehnike i računarstva, Zagreb,
Prirodoslovno-matematički fakultet, Zagreb

Poveznice na cjeloviti tekst rada:

doi www.sciencedirect.com

Citiraj ovu publikaciju:

Utrobičić, Antonija; Kovačić, Marinko; Erhardt, Filip; Poljak, Nikola; Planinić, Mirko
A floating multi-channel picoammeter for micropattern gaseous detector current monitoring // Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 801 (2015), 21-26 doi:10.1016/j.nima.2015.08.021 (međunarodna recenzija, članak, znanstveni)
Utrobičić, A., Kovačić, M., Erhardt, F., Poljak, N. & Planinić, M. (2015) A floating multi-channel picoammeter for micropattern gaseous detector current monitoring. Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 801, 21-26 doi:10.1016/j.nima.2015.08.021.
@article{article, author = {Utrobi\v{c}i\'{c}, Antonija and Kova\v{c}i\'{c}, Marinko and Erhardt, Filip and Poljak, Nikola and Planini\'{c}, Mirko}, year = {2015}, pages = {21-26}, DOI = {10.1016/j.nima.2015.08.021}, keywords = {Floating current measurement, picoammeter, digital readout, GEM detector}, journal = {Nuclear instruments and methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment}, doi = {10.1016/j.nima.2015.08.021}, volume = {801}, issn = {0168-9002}, title = {A floating multi-channel picoammeter for micropattern gaseous detector current monitoring}, keyword = {Floating current measurement, picoammeter, digital readout, GEM detector} }
@article{article, author = {Utrobi\v{c}i\'{c}, Antonija and Kova\v{c}i\'{c}, Marinko and Erhardt, Filip and Poljak, Nikola and Planini\'{c}, Mirko}, year = {2015}, pages = {21-26}, DOI = {10.1016/j.nima.2015.08.021}, keywords = {Floating current measurement, picoammeter, digital readout, GEM detector}, journal = {Nuclear instruments and methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment}, doi = {10.1016/j.nima.2015.08.021}, volume = {801}, issn = {0168-9002}, title = {A floating multi-channel picoammeter for micropattern gaseous detector current monitoring}, keyword = {Floating current measurement, picoammeter, digital readout, GEM detector} }

Časopis indeksira:


  • Current Contents Connect (CCC)
  • Web of Science Core Collection (WoSCC)
    • Science Citation Index Expanded (SCI-EXP)
    • SCI-EXP, SSCI i/ili A&HCI
  • Scopus


Citati:





    Contrast
    Increase Font
    Decrease Font
    Dyslexic Font