Pretražite po imenu i prezimenu autora, mentora, urednika, prevoditelja

Napredna pretraga

Pregled bibliografske jedinice broj: 752375

Zinc-Filled Multi-entrance Fixed Point


Zvizdić, Davor; Šestan, Danijel
Zinc-Filled Multi-entrance Fixed Point // International journal of thermophysics, 36 (2015), 2/3; 336-346 doi:10.1007/s10765-015-1846-y (međunarodna recenzija, članak, znanstveni)


CROSBI ID: 752375 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Zinc-Filled Multi-entrance Fixed Point

Autori
Zvizdić, Davor ; Šestan, Danijel

Izvornik
International journal of thermophysics (0195-928X) 36 (2015), 2/3; 336-346

Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni

Ključne riječi
Comparison calibration; Equalizing block; Multi-entrance fixed point

Sažetak
Comparison calibrations comprise by far the majority of all thermometer calibrations, and the improvement of the calibration speed and quality is of significant importance. In general, these calibrations are performed by comparing a thermometer under calibration with a reference thermometer, with both thermometers being placed in a zone of controlled temperature that is in thermal equilibrium. The time invariance of metrological characteristics of temperature regulated zones (gradients, stability, repeatability) adds a significant contribution to the total calibration uncertainty. To improve spatial thermal homogeneity and temporal stability of ordinary equalizing blocks (commonly used in combination with calibration baths and furnaces), a new concept is developed under the name multi- entrance fixed point (MEFP). With the MEFP, performances are improved by utilizing a phase transition of matter, similar to the fixed-point cells that are used for calibration of thermometers at the primary level. While fixed-point cells represent temperature standards alone, the MEFP concept requires the use of a standard thermometer as the calibration reference. This paper discusses the advantages, design possibilities, and performance of the concept and gives results of the investigation of the zinc- filled MEFP cell. The performance of the new MEFP cell was examined by calibrating a thermometer against a standard. Two calibrations were performed in the same furnace, first using an MEFP cell as an ordinary equalizing block (without a phase transition) and then at the melt plateau of zinc. The investigation results confirmed the usefulness of the MEFP concept, as the uncertainty contributions from temperature gradients and the stability of the furnace were decreased by an order of magnitude, in comparison to the ordinary equalizing block.

Izvorni jezik
Engleski

Znanstvena područja
Fizika, Strojarstvo



POVEZANOST RADA


Ustanove:
Fakultet strojarstva i brodogradnje, Zagreb

Profili:

Avatar Url Danijel Šestan (autor)

Avatar Url Davor Zvizdić (autor)

Poveznice na cjeloviti tekst rada:

doi link.springer.com

Citiraj ovu publikaciju:

Zvizdić, Davor; Šestan, Danijel
Zinc-Filled Multi-entrance Fixed Point // International journal of thermophysics, 36 (2015), 2/3; 336-346 doi:10.1007/s10765-015-1846-y (međunarodna recenzija, članak, znanstveni)
Zvizdić, D. & Šestan, D. (2015) Zinc-Filled Multi-entrance Fixed Point. International journal of thermophysics, 36 (2/3), 336-346 doi:10.1007/s10765-015-1846-y.
@article{article, author = {Zvizdi\'{c}, Davor and \v{S}estan, Danijel}, year = {2015}, pages = {336-346}, DOI = {10.1007/s10765-015-1846-y}, keywords = {Comparison calibration, Equalizing block, Multi-entrance fixed point}, journal = {International journal of thermophysics}, doi = {10.1007/s10765-015-1846-y}, volume = {36}, number = {2/3}, issn = {0195-928X}, title = {Zinc-Filled Multi-entrance Fixed Point}, keyword = {Comparison calibration, Equalizing block, Multi-entrance fixed point} }
@article{article, author = {Zvizdi\'{c}, Davor and \v{S}estan, Danijel}, year = {2015}, pages = {336-346}, DOI = {10.1007/s10765-015-1846-y}, keywords = {Comparison calibration, Equalizing block, Multi-entrance fixed point}, journal = {International journal of thermophysics}, doi = {10.1007/s10765-015-1846-y}, volume = {36}, number = {2/3}, issn = {0195-928X}, title = {Zinc-Filled Multi-entrance Fixed Point}, keyword = {Comparison calibration, Equalizing block, Multi-entrance fixed point} }

Časopis indeksira:


  • Current Contents Connect (CCC)
  • Web of Science Core Collection (WoSCC)
    • Science Citation Index Expanded (SCI-EXP)
    • SCI-EXP, SSCI i/ili A&HCI
  • Scopus


Citati:





    Contrast
    Increase Font
    Decrease Font
    Dyslexic Font