Pregled bibliografske jedinice broj: 724036
Results of external quality control of HbA1c in Croatian medical laboratories indicated that POCT system differ from immunoassay methods
Results of external quality control of HbA1c in Croatian medical laboratories indicated that POCT system differ from immunoassay methods // Biochemia Medica 24(Suppl.1) / Leniček Krleža, Jasna ; Topić, Elizabeta ; Editor-in-chief: Šimundić, Ana-Maria (ur.).
Zagreb: Medicinska naklada, 2014. str. S57-S58 (poster, međunarodna recenzija, sažetak, znanstveni)
CROSBI ID: 724036 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Results of external quality control of HbA1c in Croatian medical laboratories indicated that POCT system differ from immunoassay methods
Autori
Hrabrić Vlah, Snježana ; Antončić, Dragana ; Grzunov, Ana ; Dobrijević, Sanja ; Vlašić- Tanasković, Jelena ; Bilić-Zulle, Lidija ; Leniček Krleža, Jasna
Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, znanstveni
Izvornik
Biochemia Medica 24(Suppl.1)
/ Leniček Krleža, Jasna ; Topić, Elizabeta ; Editor-in-chief: Šimundić, Ana-Maria - Zagreb : Medicinska naklada, 2014, S57-S58
Skup
14th EFLM Countinuing Postgraduated Course in Clinical Chemistry and Laboratory Medicine
Mjesto i datum
Dubrovnik, Hrvatska, 25.10.2014. - 26.10.2014
Vrsta sudjelovanja
Poster
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
external quality control; HbA1c; interlaboratory comparability
Sažetak
Background: HbA1c as valuable marker in management of diabetes is measured in the most of medical laboratories in Croatia. Aim of this study is to present the results of external quality control assessment (EQA) in Croatian medical laboratories for HbA1c. Material and methods: Results were obtained from 93 laboratories that participated in cycle of EQA scheme in 2013.The control sample was commercial, lyophilized human blood specimen with normal HbA1c value. The results were entered on-line, as a dual system of reporting: DCCT (%) and IFCC (mmol/mol), and DCCT results were analyzed. Outliers (N=14) were excluded according to Tukey´s model and mean (X), standard deviation (SD) and coefficient of variation (CV) were calculated. According to quality specifications established in CROQALM, the allowable deviation from the mean was ±5%. Results: The results were divided in two groups according to the method (immunoassay and colorimetric method). For immunoassay methods (82 participants, six types of instruments), reported results were: X=5.4%, SD=0.18, CV=3.3%. For colorimetric method (11 participants, POCT instrument), results were: X=9.8%, SD=1.94, CV=19.8%. Immunoassay group indicated acceptable CV (<5%) while results (CV) obtained by colorimetric method (POCT analyzer) were far outside of acceptable limits. Conclusion: Immunoassay methods for HbA1c determination indicated very good inter- laboratory comparability. High CV in the group of POCT analyzer revealed poor comparability within that method group. Therefore, it can be concluded that only a minority of participants that reported results obtained on POCT analyzers demonstrated low inter-laboratory agreement.
Izvorni jezik
Engleski
Znanstvena područja
Javno zdravstvo i zdravstvena zaštita
POVEZANOST RADA
Ustanove:
Klinika za dječje bolesti Medicinskog fakulteta,
KBC "Sestre Milosrdnice",
Opća bolnica Pula,
Klinički bolnički centar Rijeka,
Klinika za dječje bolesti
Citiraj ovu publikaciju:
Časopis indeksira:
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus
- MEDLINE