Glucose determination in blood samples using flow injection analysis and an amperometric biosensor based on glucose oxidase immobilized on hexacyanoferrate modified nickel electrode


Milardović, Stjepan; Kruhak, Ivan; Iveković, Damir; Rumenjak, Vlatko; Tkalčec, Milivoj; Grabarić, Božidar S.
Glucose determination in blood samples using flow injection analysis and an amperometric biosensor based on glucose oxidase immobilized on hexacyanoferrate modified nickel electrode // Analytica chimica acta, 350 (1997), 1-2; 91-96 (međunarodna recenzija, članak, znanstveni)


CROSBI ID: 6785 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Glucose determination in blood samples using flow injection analysis and an amperometric biosensor based on glucose oxidase immobilized on hexacyanoferrate modified nickel electrode

Autori
Milardović, Stjepan ; Kruhak, Ivan ; Iveković, Damir ; Rumenjak, Vlatko ; Tkalčec, Milivoj ; Grabarić, Božidar S.

Izvornik
Analytica chimica acta (0003-2670) 350 (1997), 1-2; 91-96

Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni

Ključne riječi
amperometric glucose biosensor; clinical application; blood glucose analysis; flow injection analysis

Sažetak
A very simple amperometric enzyme-based glucose sensor is applied as detector in flow injection analysis (FIA) of diabetic patients blood glucose. The enzyme glucose oxidase (GOD) is immobilized by a glutaraldehyde/bovine serum albumin cross-linking procedure on the surface of an alkali nickel hexacyanoferrate thin film electrocrystallized on the nickel electrode. An alkali nickel hexacyanoferrate modified nickel electrode catalyzes the reduction of the hydrogen peroxide, formed by selective biocatalytic GOD action on glucose oxidation by oxygen present in sample solutions. The developed glucose sensor has shown good sensitivity at low negative operating potentials (-300 to -100 mV vs. Ag/AgCl), sufficiently wide linear glucose concentration range (5 uM - 2.9 mM), relatively long operating time (during three monthes of operation the initial sensitivity decreased only 30%) and no influence of the most common interferences present in human blood like proteins, ascorbic and uric acids.

Izvorni jezik
Engleski

Znanstvena područja
Temeljne medicinske znanosti



POVEZANOST RADA


Projekti:
108201

Ustanove:
Medicinski fakultet, Zagreb


Citiraj ovu publikaciju:

Milardović, Stjepan; Kruhak, Ivan; Iveković, Damir; Rumenjak, Vlatko; Tkalčec, Milivoj; Grabarić, Božidar S.
Glucose determination in blood samples using flow injection analysis and an amperometric biosensor based on glucose oxidase immobilized on hexacyanoferrate modified nickel electrode // Analytica chimica acta, 350 (1997), 1-2; 91-96 (međunarodna recenzija, članak, znanstveni)
Milardović, S., Kruhak, I., Iveković, D., Rumenjak, V., Tkalčec, M. & Grabarić, B. (1997) Glucose determination in blood samples using flow injection analysis and an amperometric biosensor based on glucose oxidase immobilized on hexacyanoferrate modified nickel electrode. Analytica chimica acta, 350 (1-2), 91-96.
@article{article, author = {Milardovi\'{c}, Stjepan and Kruhak, Ivan and Ivekovi\'{c}, Damir and Rumenjak, Vlatko and Tkal\v{c}ec, Milivoj and Grabari\'{c}, Bo\v{z}idar S.}, year = {1997}, pages = {91-96}, keywords = {amperometric glucose biosensor, clinical application, blood glucose analysis, flow injection analysis}, journal = {Analytica chimica acta}, volume = {350}, number = {1-2}, issn = {0003-2670}, title = {Glucose determination in blood samples using flow injection analysis and an amperometric biosensor based on glucose oxidase immobilized on hexacyanoferrate modified nickel electrode}, keyword = {amperometric glucose biosensor, clinical application, blood glucose analysis, flow injection analysis} }
@article{article, author = {Milardovi\'{c}, Stjepan and Kruhak, Ivan and Ivekovi\'{c}, Damir and Rumenjak, Vlatko and Tkal\v{c}ec, Milivoj and Grabari\'{c}, Bo\v{z}idar S.}, year = {1997}, pages = {91-96}, keywords = {amperometric glucose biosensor, clinical application, blood glucose analysis, flow injection analysis}, journal = {Analytica chimica acta}, volume = {350}, number = {1-2}, issn = {0003-2670}, title = {Glucose determination in blood samples using flow injection analysis and an amperometric biosensor based on glucose oxidase immobilized on hexacyanoferrate modified nickel electrode}, keyword = {amperometric glucose biosensor, clinical application, blood glucose analysis, flow injection analysis} }

Časopis indeksira:


  • Current Contents Connect (CCC)
  • Web of Science Core Collection (WoSCC)
    • Science Citation Index Expanded (SCI-EXP)
    • SCI-EXP, SSCI i/ili A&HCI
  • Scopus





Contrast
Increase Font
Decrease Font
Dyslexic Font