A New Two Temperature Humidity Generator as a Primary Standard For Measuring Humidity in Croatia


Zvizdić, Davor; Kolobarić, Vladimir; Galaso, Ivan; Šerfezi, Damir
A New Two Temperature Humidity Generator as a Primary Standard For Measuring Humidity in Croatia // TEMPMEKO '99 Proceedings I / Dubbeldam, Jaco F. ; de Groot, Martin J. (ur.).
Delft: NMi Van Swinden Laboratorium, 1999. str. 203-208 (poster, međunarodna recenzija, cjeloviti rad (in extenso), znanstveni)


CROSBI ID: 65285 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
A New Two Temperature Humidity Generator as a Primary Standard For Measuring Humidity in Croatia

Autori
Zvizdić, Davor ; Kolobarić, Vladimir ; Galaso, Ivan ; Šerfezi, Damir

Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni

Izvornik
TEMPMEKO '99 Proceedings I / Dubbeldam, Jaco F. ; de Groot, Martin J. - Delft : NMi Van Swinden Laboratorium, 1999, 203-208

Skup
The 7th International Symposium on Temperature and Thermal Measurements in Industry and Science TEMPMEKO '99

Mjesto i datum
Delft, Nizozemska, 01.06.1999. - 03.06.1999

Vrsta sudjelovanja
Poster

Vrsta recenzije
Međunarodna recenzija

Ključne riječi
humidity standard; calibration; uncertainty

Sažetak
This paper describes the principle and design of two temperature humidity generator being developed at the Laboratory for Process Measurement. It is designed to eventually operate in dew point range from -15°C to 60°C and in test chamber range to 90°C, thus catering to most of the present demand for calibration in Croatia. The flow through saturator, test chamber and test chamber bypass loop is monitored and set to the desired levels with flow meters and valves. It can be adjusted from fully recirculating flow to fully open circuit flow. Gas saturation degree is determined by measuring the temperature differences between saturator gas outlet and liquid at the saturator inlet. Temperature controllers, thermometers and resistance bridge are connected to computer via digital interfaces and integrated into monitoring and control system on the custom designed, Labview based platform. The initial results of generator evaluation are presented for recirculating mode and for part of the operating range. They include saturator and test chamber responses to stepwise parameter changes and uncertainty estimates in terms of dew point and relative humidity.

Izvorni jezik
Engleski

Znanstvena područja
Strojarstvo



POVEZANOST RADA


Projekti:
120007
120043

Ustanove:
Fakultet strojarstva i brodogradnje, Zagreb

Profili:

Avatar Url Ivan Galaso (autor)

Avatar Url Damir Šerfezi (autor)

Avatar Url Davor Zvizdić (autor)

Avatar Url Vladimir Kolobarić (autor)


Citiraj ovu publikaciju:

Zvizdić, Davor; Kolobarić, Vladimir; Galaso, Ivan; Šerfezi, Damir
A New Two Temperature Humidity Generator as a Primary Standard For Measuring Humidity in Croatia // TEMPMEKO '99 Proceedings I / Dubbeldam, Jaco F. ; de Groot, Martin J. (ur.).
Delft: NMi Van Swinden Laboratorium, 1999. str. 203-208 (poster, međunarodna recenzija, cjeloviti rad (in extenso), znanstveni)
Zvizdić, D., Kolobarić, V., Galaso, I. & Šerfezi, D. (1999) A New Two Temperature Humidity Generator as a Primary Standard For Measuring Humidity in Croatia. U: Dubbeldam, J. & de Groot, M. (ur.)TEMPMEKO '99 Proceedings I.
@article{article, author = {Zvizdi\'{c}, Davor and Kolobari\'{c}, Vladimir and Galaso, Ivan and \v{S}erfezi, Damir}, year = {1999}, pages = {203-208}, keywords = {humidity standard, calibration, uncertainty}, title = {A New Two Temperature Humidity Generator as a Primary Standard For Measuring Humidity in Croatia}, keyword = {humidity standard, calibration, uncertainty}, publisher = {NMi Van Swinden Laboratorium}, publisherplace = {Delft, Nizozemska} }
@article{article, author = {Zvizdi\'{c}, Davor and Kolobari\'{c}, Vladimir and Galaso, Ivan and \v{S}erfezi, Damir}, year = {1999}, pages = {203-208}, keywords = {humidity standard, calibration, uncertainty}, title = {A New Two Temperature Humidity Generator as a Primary Standard For Measuring Humidity in Croatia}, keyword = {humidity standard, calibration, uncertainty}, publisher = {NMi Van Swinden Laboratorium}, publisherplace = {Delft, Nizozemska} }




Contrast
Increase Font
Decrease Font
Dyslexic Font