Verification of claimed limit of detection for the Roche Troponin T hs STAT assay
Verification of claimed limit of detection for the Roche Troponin T hs STAT assay // 2nd EFLM-BD European Conference on Preanalytical Phase "Preanalytical quality improvement – in quality we trust" : abstracts ; u: Biochemia Medica / Šimundić, Ana-Maria (ur.).
Zagreb: Medicinska naklada, 2013. str. A38-A39 (poster, nije recenziran, sažetak, znanstveni)
CROSBI ID: 650740 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Verification of claimed limit of detection for the Roche Troponin T hs STAT assay
Autori
Dukić, Lora ; Šimundić, Ana-Maria ; Malogorski, Davorin
Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, znanstveni
Izvornik
2nd EFLM-BD European Conference on Preanalytical Phase "Preanalytical quality improvement – in quality we trust" : abstracts ; u: Biochemia Medica
/ Šimundić, Ana-Maria - Zagreb : Medicinska naklada, 2013, A38-A39
Skup
EFLM-BD European Conference on Preanalytical Phase "Preanalytical quality improvement – in quality we trust" (2 ; 2013)
Mjesto i datum
Zagreb, Hrvatska, 01.03.2013. - 02.03.2013
Vrsta sudjelovanja
Poster
Vrsta recenzije
Nije recenziran
Ključne riječi
limit of detection ; verification ; troponin T assay
Sažetak
Detection of cardiac biomarker troponin T (cTnT) together with other evidence of myocardial ischaemia makes important criteria for the diagnosis of myocardial infarction (MI). Limit of detection (LoD) is defined as the lowest amount of analyte in a sample that can be detected by given analytical procedure. Establishment of the limit of detection for cTnT has important role in assessment of patient suspected for MI. The aim of our study was to verify the claimed limit of detection of 0.005 μg/L for the cTnT assay (Roche Troponin T hs (high sensitive) STAT (Short Turn Around Time)) according the Clinical and Laboratory Standards Institute (CLSI) EP17-A protocol. Materials and methods: Verification of the claimed LoD was performed on Cobas e 411 analyzer (Roche Diagnostics GmbH, Mannheim, Germany). Twenty repeated measurements of pooled plasma samples with concentration of cTnT 0.005 μg/L were performed. Results: Claimed LoD for the Roche Troponin T hs STAT assay showed that observed proportion (20/20) is in accordance with the expected value (95%) at concentration of 0.005 μg/L. Conclusions: Our data support the LoD claimed by Roche for Troponin T hs STAT assay.
Izvorni jezik
Engleski
Znanstvena područja
Kliničke medicinske znanosti
POVEZANOST RADA
Projekti:
134-1340227-0200 - Upala i udio farmakogenetike u razvoju i ishodu akutnih i kroničnih bolesti (Šimundić, Ana-Maria, MZOS ) ( CroRIS)
Ustanove:
KBC "Sestre Milosrdnice"
Citiraj ovu publikaciju:
Časopis indeksira:
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus
- MEDLINE