PRIMARY DEW-POINT GENERATION BETWEEN 1°C AND 60°C AT LPM


Zvizdic, Davor; Heinonen, Martti; Sestan, Danijel
PRIMARY DEW-POINT GENERATION BETWEEN 1°C AND 60°C AT LPM // 20th IMEKO World Congress 2012
Red Hook (NY): Curran Associates, 2013. (predavanje, nije recenziran, cjeloviti rad (in extenso), znanstveni)


CROSBI ID: 649242 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
PRIMARY DEW-POINT GENERATION BETWEEN 1°C AND 60°C AT LPM

Autori
Zvizdic, Davor ; Heinonen, Martti ; Sestan, Danijel

Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni

Izvornik
20th IMEKO World Congress 2012 / - Red Hook (NY) : Curran Associates, 2013

ISBN
978-1-62748-190-8

Skup
XX IMEKO World Congress

Mjesto i datum
Busan, Republika Koreja, 09.09.2012. - 14.09.2012

Vrsta sudjelovanja
Predavanje

Vrsta recenzije
Nije recenziran

Ključne riječi
dew-point generator; humidity; saturator

Sažetak
EUROMET collaborative project has been set up between Finish Centre for Metrology and Accreditation (MIKES) and Croatian Laboratory for process measurements (HMI/FSB-LPM) with the objective to provide HMI/FSB-LPM with two new primary dew-point generators. Both generators operate on single pressure – single pass principle and together cover the dew-point temperature range from -70°C to 60°C, with gas flow rates up to 2.6L/min. This article presents the design of the high range generator (HRS), which operates in a dew-point range between 1°C and 60°C. To investigate how close the generator is to the ideal saturation, MIKES conducted several efficiency tests, which will also be described. Results of the tests show that HRS generator is sufficiently efficient for a primary realization of the dew-point temperature scale in the specified ranges. The estimated standard uncertainties due to the non-ideal saturation efficiency are found to be between 0.02 °C and 0.05 °C

Izvorni jezik
Engleski

Znanstvena područja
Strojarstvo



POVEZANOST RADA


Projekti:
120-0000000-3322 - Razvoj nacionalnog etalona temperature (Zvizdić, Davor, MZOS ) ( CroRIS)

Ustanove:
Fakultet strojarstva i brodogradnje, Zagreb

Profili:

Avatar Url Danijel Šestan (autor)

Avatar Url Davor Zvizdić (autor)

Poveznice na cjeloviti tekst rada:

Pristup cjelovitom tekstu rada

Citiraj ovu publikaciju:

Zvizdic, Davor; Heinonen, Martti; Sestan, Danijel
PRIMARY DEW-POINT GENERATION BETWEEN 1°C AND 60°C AT LPM // 20th IMEKO World Congress 2012
Red Hook (NY): Curran Associates, 2013. (predavanje, nije recenziran, cjeloviti rad (in extenso), znanstveni)
Zvizdic, D., Heinonen, M. & Sestan, D. (2013) PRIMARY DEW-POINT GENERATION BETWEEN 1°C AND 60°C AT LPM. U: 20th IMEKO World Congress 2012.
@article{article, author = {Zvizdic, Davor and Heinonen, Martti and Sestan, Danijel}, year = {2013}, keywords = {dew-point generator, humidity, saturator}, isbn = {978-1-62748-190-8}, title = {PRIMARY DEW-POINT GENERATION BETWEEN 1°C AND 60°C AT LPM}, keyword = {dew-point generator, humidity, saturator}, publisher = {Curran Associates}, publisherplace = {Busan, Republika Koreja} }
@article{article, author = {Zvizdic, Davor and Heinonen, Martti and Sestan, Danijel}, year = {2013}, keywords = {dew-point generator, humidity, saturator}, isbn = {978-1-62748-190-8}, title = {PRIMARY DEW-POINT GENERATION BETWEEN 1°C AND 60°C AT LPM}, keyword = {dew-point generator, humidity, saturator}, publisher = {Curran Associates}, publisherplace = {Busan, Republika Koreja} }




Contrast
Increase Font
Decrease Font
Dyslexic Font