Comparison of Haemoglobin and Haematocrit Levels On Point of Care Blood Gas Analyzer and Automated Haematology Analyzer
Comparison of Haemoglobin and Haematocrit Levels On Point of Care Blood Gas Analyzer and Automated Haematology Analyzer // 13th EFLM Continuous Postgraduate Course in Clinical Chemistry and Laboratory Medicine: New Trends in Diagnosis and Monitoring using POC Instruments / Biochemia Medica, Šimundić, AM (ur.).
Zagreb: Medicinska naklada, 2013. str. A83-A83 (poster, međunarodna recenzija, sažetak, znanstveni)
CROSBI ID: 648715 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Comparison of Haemoglobin and Haematocrit Levels On Point of Care Blood Gas Analyzer and Automated Haematology Analyzer
Autori
Grzunov, Ana ; Leniček Krleža, Jasna
Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, znanstveni
Izvornik
13th EFLM Continuous Postgraduate Course in Clinical Chemistry and Laboratory Medicine: New Trends in Diagnosis and Monitoring using POC Instruments
/ Biochemia Medica, Šimundić, AM - Zagreb : Medicinska naklada, 2013, A83-A83
Skup
13th EFLM Continuous Postgraduate Course in Clinical Chemistry and Laboratory Medicine
Mjesto i datum
Dubrovnik, Hrvatska, 19.10.2013. - 20.10.2013
Vrsta sudjelovanja
Poster
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
point–of-care; blood gas analyzer; haematology analyzer; haemoglobin; haematocrit
Sažetak
Background: The aim of this study was to compare haemoglobin and haematocrit levels measured as point of care testing and measurements done by automated haematology analyzer. Materials and methods: In this prospective study, haemoglobin and haematocrit levels were measured in 126 consecutive paired arterial or capillary samples using a point of care blood gas analyzer and automated haematology analyzer. Pearson correlation coefficient and Passing-Bablok regression were calculated for the data comparison. Results: Mean (SD) level of haemoglobin measured on blood gas analyzer was 135.59 g/L (26.48) and measured on automated haematology analyzer 118.84 g/L (25.59). The mean difference between the blood gas analyzer and automated haematology analyzer was -16.75 g/L (P < 0.001). Mean (SD) level of haematocrit measured on blood gas analyzer was 0.399 L/L (0.078) and measured on automated haematology analyzer 0.356 L/L (0.071), respectively. The mean difference between the blood gas analyzer and automated haematology analyzer was -0.043 L/L (P < 0.001). Comparison of the results of two devices yielded correlation coefficients r = 0.98 (P < 0.001) for haemoglobin and r = 0.75 (P < 0.001) for haematocrit. Conclusions: Although the correlations between the measurements were significant, blood gas analyzer has shown unacceptable agreement with automated haematology analyzer. It overestimates haemoglobin and haematocrit values, which is not clinically acceptable.
Izvorni jezik
Engleski
Znanstvena područja
Kliničke medicinske znanosti
POVEZANOST RADA
Projekti:
098-0982464-3120 - Farmakogenetika u dječjoj onkologiji (Stepan Giljević, Jasminka, MZOS ) ( CroRIS)
Ustanove:
Klinika za dječje bolesti Medicinskog fakulteta
Profili:
Jasna Leniček Krleža
(autor)
Citiraj ovu publikaciju:
Časopis indeksira:
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus
- MEDLINE