Pregled bibliografske jedinice broj: 64839
Investigation of Dynamic Strength of Ag-Pd Alloy and Hydrothermal ceramics
Investigation of Dynamic Strength of Ag-Pd Alloy and Hydrothermal ceramics // 22nd Annual Conference "Novel biomaterials in prosthodonics and implantology and clinically relevant aspects on occlusion" / - (ur.).
Turku, Finska: -, 1998. str. 78-78 (poster, međunarodna recenzija, sažetak, znanstveni)
CROSBI ID: 64839 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Investigation of Dynamic Strength of Ag-Pd Alloy and Hydrothermal ceramics
Autori
Poljak-Guberina, Renata ; Ćatović, Adnan ; Jerolimov, Vjekoslav ; Franz, Mladen
Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, znanstveni
Izvornik
22nd Annual Conference "Novel biomaterials in prosthodonics and implantology and clinically relevant aspects on occlusion"
/ - , 1998, 78-78
Skup
22nd Annual Conference "Novel biomaterials in prosthodonics and implantology and clinically relevant aspects on occlusion"
Mjesto i datum
Turku, Finska, 27.08.1998. - 29.08.1998
Vrsta sudjelovanja
Poster
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
Fatigue strength; Ag-Pd alloy; hydrothermal ceramic
Sažetak
The aim of study:
1) Determine fatigue strength of interface between Ag-Pd alloy and hydrothermal ceramics.
2) Determine the inluence of thermal cycling on the fatigue strength of tested metalceramic system.
Material and methods:
Investigations were performed on 48 Ag-Pd alloy samples (45x12x4 mm)on which 2 mm thick hydrothermal ceramics was fused in length of 25 mm. The first group of 24 samples was dynamically loaded immediately after casting and fusing. The second group was thermocycled in the water 1000 times at temperatures of 5 C and 55 C before dynamic loading. The three-point dynamic loading test was performed in Amsler's high-frequency pulsator within the loading range from 550 N to 1200 N. The dinamic force values and number of cycles were measured until the first fracture occured between metal and ceramics.
Conclusions:
1) Endurance limit for nonthermocycled sample is 850 N.
2) Endurance limit for thermocycled sample is 550 N.
3) Thermocycling reduces the fatigue strength of the Ag-Pd alloy and hydrothermal ceramics system for 28 %.