Intercomparison of the Dew-Point Temperature Realizations at LPM and MIKES in the Range from −70 ◦C to +20 ◦C


Heinonen, Martti; Zvizdić, Davor; Šestan, Danijel
Intercomparison of the Dew-Point Temperature Realizations at LPM and MIKES in the Range from −70 ◦C to +20 ◦C // International journal of thermophysics, 33 (2011), 8/9; 1451-1457 doi:10.1007/s10765-011-1105-9 (međunarodna recenzija, članak, znanstveni)


CROSBI ID: 611055 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Intercomparison of the Dew-Point Temperature Realizations at LPM and MIKES in the Range from −70 ◦C to +20 ◦C

Autori
Heinonen, Martti ; Zvizdić, Davor ; Šestan, Danijel

Izvornik
International journal of thermophysics (0195-928X) 33 (2011), 8/9; 1451-1457

Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni

Ključne riječi
bilateral degrees of equivalence; Dew-point temperature; supplementary comparison

Sažetak
The first European humidity key comparison EURAMET-T.K6 was completed in 2008, and it covered the dew-point temperature range from −50 ◦C to +20 ◦C. Both LPM and MIKES participated in the comparison, but a new low dew-point generator was introduced at LPM as a result of progress in the EUROMET P912 project. To extend the range of available comparison evidence down to −70 ◦C and to study the validity of improved uncertainties of LPM, a bilateral comparison was carried out between LPM and MIKES in 2009–2010. The applied comparison procedure was similar to that applied in EURAMET-T.K6. However, only one transfer standard was used instead of two units and the measurement point −70 ◦C was added in the measurement scheme. The results show that the bilateral equivalence between LPM and MIKES is between (0.00±0.06) ◦C and (0.02±0.08) ◦C in the range from −50 ◦C to +20 ◦C and (0.01 ± 0.10) ◦C at −70 ◦C. Using MIKES results as the link to the EURAMET.T-K6, it is shown that the difference between the results obtained with the new LPM dew- point temperature standard and the EURAMET Comparison Reference Values is between (−0.02 ± 0.08) ◦C at 20◦C and (+0.02 ± 0.07) ◦C at −50 ◦C.

Izvorni jezik
Engleski

Znanstvena područja
Strojarstvo



POVEZANOST RADA


Projekti:
120-0000000-3322 - Razvoj nacionalnog etalona temperature (Zvizdić, Davor, MZOS ) ( CroRIS)

Ustanove:
Fakultet strojarstva i brodogradnje, Zagreb

Profili:

Avatar Url Danijel Šestan (autor)

Avatar Url Davor Zvizdić (autor)

Poveznice na cjeloviti tekst rada:

doi link.springer.com

Citiraj ovu publikaciju:

Heinonen, Martti; Zvizdić, Davor; Šestan, Danijel
Intercomparison of the Dew-Point Temperature Realizations at LPM and MIKES in the Range from −70 ◦C to +20 ◦C // International journal of thermophysics, 33 (2011), 8/9; 1451-1457 doi:10.1007/s10765-011-1105-9 (međunarodna recenzija, članak, znanstveni)
Heinonen, M., Zvizdić, D. & Šestan, D. (2011) Intercomparison of the Dew-Point Temperature Realizations at LPM and MIKES in the Range from −70 ◦C to +20 ◦C. International journal of thermophysics, 33 (8/9), 1451-1457 doi:10.1007/s10765-011-1105-9.
@article{article, author = {Heinonen, Martti and Zvizdi\'{c}, Davor and \v{S}estan, Danijel}, year = {2011}, pages = {1451-1457}, DOI = {10.1007/s10765-011-1105-9}, keywords = {bilateral degrees of equivalence, Dew-point temperature, supplementary comparison}, journal = {International journal of thermophysics}, doi = {10.1007/s10765-011-1105-9}, volume = {33}, number = {8/9}, issn = {0195-928X}, title = {Intercomparison of the Dew-Point Temperature Realizations at LPM and MIKES in the Range from −70 ◦C to +20 ◦C}, keyword = {bilateral degrees of equivalence, Dew-point temperature, supplementary comparison} }
@article{article, author = {Heinonen, Martti and Zvizdi\'{c}, Davor and \v{S}estan, Danijel}, year = {2011}, pages = {1451-1457}, DOI = {10.1007/s10765-011-1105-9}, keywords = {bilateral degrees of equivalence, Dew-point temperature, supplementary comparison}, journal = {International journal of thermophysics}, doi = {10.1007/s10765-011-1105-9}, volume = {33}, number = {8/9}, issn = {0195-928X}, title = {Intercomparison of the Dew-Point Temperature Realizations at LPM and MIKES in the Range from −70 ◦C to +20 ◦C}, keyword = {bilateral degrees of equivalence, Dew-point temperature, supplementary comparison} }