Pregled bibliografske jedinice broj: 551282
Use of high-resolution digital multimeters for the calibration of temperature sensors
Use of high-resolution digital multimeters for the calibration of temperature sensors // Proceeding of the 17th Symposium IMEKO TC 4, 3rd Symposium IMEKO TC 19 and 15th IWADC Workshop / Michaeli, Linus ; Šaliga, Jan (ur.).
Košice: International Measurement Confederation (IMEKO), 2010. str. 1-5 (predavanje, međunarodna recenzija, cjeloviti rad (in extenso), znanstveni)
CROSBI ID: 551282 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Use of high-resolution digital multimeters for the calibration of temperature sensors
Autori
Ilić, Damir ; Leniček, Ivan ; Ferković, Luka
Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni
Izvornik
Proceeding of the 17th Symposium IMEKO TC 4, 3rd Symposium IMEKO TC 19 and 15th IWADC Workshop
/ Michaeli, Linus ; Šaliga, Jan - Košice : International Measurement Confederation (IMEKO), 2010, 1-5
ISBN
978-80-553-0424-3
Skup
17th Symposium IMEKO TC 4 (17 ; 2010) ; Symposium IMEKO TC 19 (3 ; 2008) ; IWADC Workshop (15 ; 2008)
Mjesto i datum
Košice, Slovačka, 08.09.2010. - 10.09.2010
Vrsta sudjelovanja
Predavanje
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
NTCRs; digital multimeter; Pt-25; calibration
Sažetak
The calibration of precise negative temperature coefficient resistors (NTCRs), which have resistance RN = 10000 ohm at a temperature of 25 °C and limits of error 0, 2 °C in the range from 0 °C to 70 °C, by means of the reference platinum resistance thermometer Pt-25 (with the resistance of ≈25, 5 ohm at 0 °C) in the range of (23 +- 4) °C, and the high-resolution digital multimeters (up to 8½ digits), is described. The influence of the resolution of the digital multimeter is investigated, as well as the influence of the self-heating of NTCRs and the measurement on different ranges and with different currents during the comparison of different type of sensors. The analysis of the resistance-versus-temperature (R-T) dependence is done by using the least-squares method and the two-parameter approximation curve for NTCRs. The results have shown that, in the range of (23 +- 4) °C, which is useful for the laboratory applications, NTCRs can be calibrated by means of Pt-25 thermometer and high-resolution digital multimeters with an uncertainty of a few millikelvin, which is very convenient for their wide application as precise and sensitive temperature sensors.
Izvorni jezik
Engleski
Znanstvena područja
Elektrotehnika
POVEZANOST RADA
Projekti:
036-0363081-1650 - Mjerenje elektromagnetskih veličina na etalonskoj razini (Ilić, Damir, MZO ) ( CroRIS)
Ustanove:
Fakultet elektrotehnike i računarstva, Zagreb