Evaluation of crossfloat measurements with pressure balances – Results of EURAMET project 1125
Evaluation of crossfloat measurements with pressure balances – Results of EURAMET project 1125 // PTB Mitteilungen, 3 (2011), 294-297 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 533097 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Evaluation of crossfloat measurements with pressure balances – Results of EURAMET project 1125
Autori
P. Otal, I. Morgado, D. Steindl, N. Medina Martín, A. Lefkopoulos, A. Altintas, L. Grgec Bermanec, S. Bursic, J. Setina, M. Bergoglio, I. Spohr, C. Wüthrich, M. Rantanen, C. Vámossy1, W. Sabuga, I. Koçaş
Izvornik
PTB Mitteilungen (0030-834X) 3
(2011);
294-297
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
pressure balance; effective area; crossfloat; pressure distortion coefficient
Sažetak
Fifteen European metrological institutes have compared their approach to estimate the effective area parameters resulting from the calibration of pressure balances by crossfloating . The methods in estimating these parameters and the associated uncertainties are tested by applying four calibration data sets numerically simulated, two in gas (up to 10 MPa) and two in oil (up to 500 MPa). The performance of the methods is evaluated in terms of robustness and ability to estimate the uncertainties of the effective area parameters.
Izvorni jezik
Engleski
Znanstvena područja
Strojarstvo
POVEZANOST RADA
Projekti:
120-0000000-3322 - Razvoj nacionalnog etalona temperature (Zvizdić, Davor, MZOS ) ( CroRIS)
Ustanove:
Fakultet strojarstva i brodogradnje, Zagreb
Profili:
Lovorka Grgec-Bermanec
(autor)
Citiraj ovu publikaciju:
Časopis indeksira:
- Scopus