Pregled bibliografske jedinice broj: 517845
Design, instrumentation, and the light response of the Heavy-Ion detector array inside the Big-Bite Spectrometer
Design, instrumentation, and the light response of the Heavy-Ion detector array inside the Big-Bite Spectrometer // Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 631 (2011), 1; 62-67 doi:10.1016/j.nima.2010.12.099 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 517845 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Design, instrumentation, and the light response of the Heavy-Ion detector array inside the Big-Bite Spectrometer
Autori
Gašparić, Igor ; Joulaeizadeh, L. ; Bacelar, J.C.S. ; van den Berg, A.M. ; Čaplar, Roman ; Loehner, H.
Izvornik
Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment (0168-9002) 631
(2011), 1;
62-67
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
CsI(Tl) scintillator ; Phoswich Light ; output particle identification ; Magnetic spectrometer
Sažetak
A phoswich array consisting of 60 elements has been built and used in pionic-fusion experiments at the Kernfysisch Versneller Instituut (KVI). The array is used in conjunction with a magnetic spectrometer and a time-of-flight system to measure the fragments produced in 3He+4He and 4He+6Li reactions at beam energies of 85.3 and 59.1 A MeV, respectively. The particle identification for Z≤5 has been performed by means of time-of-flight and pulse-shape analysis techniques. A momentum resolution of 1% was achieved. The dependence of the integrated light output on ion energy, momentum, and mass has been studied. The detector performance, light response, ion identification, momentum and time calibration, and acceptance of the Heavy-Ion detector array are presented and discussed.
Izvorni jezik
Engleski
Znanstvena područja
Fizika
POVEZANOST RADA
Projekti:
098-1191005-2879 - Fizika teških iona (Basrak, Zoran, MZOS ) ( CroRIS)
Ustanove:
Institut "Ruđer Bošković", Zagreb
Citiraj ovu publikaciju:
Časopis indeksira:
- Current Contents Connect (CCC)
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus