A Novel Approach to Modeling Coronary Stents Using a Slender Curved Rod Model: A Comparison Between Fractured Xience-Like and Palmaz-Like Stents
A Novel Approach to Modeling Coronary Stents Using a Slender Curved Rod Model: A Comparison Between Fractured Xience-Like and Palmaz-Like Stents // Applied and Numerical Partial Differential Equations / Fitzgibbon, William ; Kuznetsov, Yuri ; Neittaanmäki, Pekka ; Périaux, Jacques ; Pironneau, Olivier (ur.).
Dordrecht : Heidelberg : New York (NY) : London: Springer, 2010. str. 41-58
CROSBI ID: 446831 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
A Novel Approach to Modeling Coronary Stents Using a Slender Curved Rod Model: A Comparison Between Fractured Xience-Like and Palmaz-Like Stents
Autori
Tambača, Josip ; Čanić, Sunčica ; Paniagua, David
Vrsta, podvrsta i kategorija rada
Poglavlja u knjigama, znanstveni
Knjiga
Applied and Numerical Partial Differential Equations
Urednik/ci
Fitzgibbon, William ; Kuznetsov, Yuri ; Neittaanmäki, Pekka ; Périaux, Jacques ; Pironneau, Olivier
Izdavač
Springer
Grad
Dordrecht : Heidelberg : New York (NY) : London
Godina
2010
Raspon stranica
41-58
ISBN
978-90-481-3238-6
Ključne riječi
endovascular stents, curved rod, fractured stents, Palmaz-like stent, Xience-like stent
Sažetak
We present a novel mathematical model to study the mechanical properties of endovascular stents in their expanded state. The model is based on the theory of slender curved rods. Stent struts are modeled as linearly elastic curved rods that satisfy the kinematic and dynamic contact conditions at the vertices where the struts meet. A weak formulation for the stent problem is defined and a Finite Element Method for a numerical computation of its solution is used to study mechanical properties of two commonly used coronary stents (Palmaz-like and Xience-like stent) in their expanded, fractured state. A simple fracture (separation), corresponding to one stent strut being disconnected from one vertex in a stent, was considered. Our results show a drastic difference in the response of the two stents to the physiologically reasonable uniform compression and bending forces.
Izvorni jezik
Engleski
Znanstvena područja
Matematika
POVEZANOST RADA
Projekti:
037-0693014-2765 - Matematička analiza kompozitnih i tankih struktura (Tutek, Zvonimir, MZOS ) ( CroRIS)
Ustanove:
Prirodoslovno-matematički fakultet, Matematički odjel, Zagreb
Profili: