Pregled bibliografske jedinice broj: 377068
The use of the “ ALARA” dosimeter in personal dose assessment in nuclear medicine procedures
The use of the “ ALARA” dosimeter in personal dose assessment in nuclear medicine procedures // Book of abstracts, Third international congress of the Croatian Society of Nuclear Medicine
Zagreb, 1999. (predavanje, međunarodna recenzija, sažetak, znanstveni)
CROSBI ID: 377068 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
The use of the “ ALARA” dosimeter in personal dose assessment in nuclear medicine procedures
Autori
Bokulić, Tomislav ; Budanec, Mirjana ; Vekić, Branko ; Lacić, Miodrag ; Kusić, Zvonko
Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, znanstveni
Izvornik
Book of abstracts, Third international congress of the Croatian Society of Nuclear Medicine
/ - Zagreb, 1999
Skup
Third international congress of the Croatian Society of Nuclear Medicine
Mjesto i datum
Opatija, Hrvatska, 10.10.1999. - 22.10.1999
Vrsta sudjelovanja
Predavanje
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
personal dosimeter; ALARA
Sažetak
Personal radiation doses have been measured with electronic dosimeters (ALARA, IMI) in typical nuclear medicine procedures. The characterisation of dosimeters and energy response was first estimated with Ra-226, Cs-137 and Tc-99m point sources. In personal dose monitoring programme, the most exposed group (mean 33.2 microSv/day) was the group of technicians preparing the individual doses and performing bone and thyroid scans. Much lower doses were observed in another group of technicians who conducted brain scintigraphy, ventriculography and immunoscintigraphy. In remaining groups the doses recorded were at the level of 3-4 microSv/day.
Izvorni jezik
Engleski
Znanstvena područja
Fizika, Kliničke medicinske znanosti
POVEZANOST RADA
Ustanove:
KBC "Sestre Milosrdnice",
Zdravstveno veleučilište, Zagreb
Profili:
Mirjana Budanec
(autor)
Zvonko Kusić
(autor)
Miodrag Lacić
(autor)
Tomislav Bokulić
(autor)
Branko Vekić
(autor)