LET dependence of high sensitivity TL dosemeters
LET dependence of high sensitivity TL dosemeters // Radiation Protection Dosimetry, 84 (1999), 1-4; 219-222 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 36886 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
LET dependence of high sensitivity TL dosemeters
Autori
Osvay, Margit ; Ranogajec-Komor, Maria
Izvornik
Radiation Protection Dosimetry (0144-8420) 84
(1999), 1-4;
219-222
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Sažetak
LET dependence of highly sensitive Al2O3:C and CaF2:Mn TL dosimeters was investigated using glow curve analysis to determine the efficiencies of TL light production for alpha particle and neutron irradiation relative to gamma exposure. No high temperature peak, nor significant change of the glow curve shape were found after irradiation by high LET (alpha and neutron) radiation. Gamma sensitivity of Al2O3:C as compared to CaF2:Mn was 6:1. However the relative TL efficiency for alpha particle irradiation of Al2O3 was a factor ten lower as compared to that of CaF2:Mn. Neutron sensitivity was found to be similar for the two materials.