Analytical performance of Accu Chek Compact Plus blood glucometer
Analytical performance of Accu Chek Compact Plus blood glucometer // Clinical Chemistry and Laboratory Medicine / Siest, Gerard (ur.).
Berlin: Walter de Gruyter, 2008. str. A79-A79 (poster, međunarodna recenzija, sažetak, stručni)
CROSBI ID: 363752 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Analytical performance of Accu Chek Compact Plus blood glucometer
Autori
Šimundić, Ana-Maria ; Nikolac, Nora ; Kozar, Jelena ; Topić, Elizabeta
Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, stručni
Izvornik
Clinical Chemistry and Laboratory Medicine
/ Siest, Gerard - Berlin : Walter de Gruyter, 2008, A79-A79
Skup
Swiss MedLab International Congress
Mjesto i datum
Montreux, Švicarska, 16.09.2008. - 19.09.2008
Vrsta sudjelovanja
Poster
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
blood glucometer; analytical performance
Sažetak
Objectives: The aim of this study was to evaluate the analytical performance characteristics of the Accu Chek Compact Plus blood glucometer by comparing its accuracy and precision with our current laboratory reference analytical system – Olympus 2700 automatic analyser. Materials and methods: Original Olympus glucose reagent served as a reference method. The difference between glucose concentrations measured on reference and Accu Chek analytical platform was tested by paired t-test. Glucometer accuracy was evaluated using bias (%), Passing-Bablock regression, Bland-Altman and Error Grid Clark analysis. Intra-assay glucometer precision was examined by 10 consecutive measurements at two commercially available control samples and one random patient sample. Results: The average Accu Chek bias was -6.6% ; we found no significant deviation from linearity ; more than 95% of our values were between mean± ; ; 1.96 SD. The results of the Clarke Error Grid analysis were 100% in zone A. Conclusion: Accu Chek Compact Plus blood glucometer has satisfying accuracy and acceptable precision and may, therefore, be used interchangeably with our laboratory referent-system for glucose measurement.
Izvorni jezik
Engleski
Znanstvena područja
Kliničke medicinske znanosti
POVEZANOST RADA
Projekti:
134-1340227-0200 - Upala i udio farmakogenetike u razvoju i ishodu akutnih i kroničnih bolesti (Šimundić, Ana-Maria, MZOS ) ( CroRIS)
Ustanove:
KBC "Sestre Milosrdnice"
Citiraj ovu publikaciju:
Časopis indeksira:
- Current Contents Connect (CCC)
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus
- MEDLINE