Pretražite po imenu i prezimenu autora, mentora, urednika, prevoditelja

Napredna pretraga

Pregled bibliografske jedinice broj: 348002

Determination of cationic surfactants in pharmaceutical disinfectants using a new, highly sensitive potentiometric sensor


Madunić-Čačić, Dubravka; Sak-Bosnar, Milan; Galović, Olivera; Sakač, Nikola; Matešić-Puač, Ružica; Grabarić, Zorana
Determination of cationic surfactants in pharmaceutical disinfectants using a new, highly sensitive potentiometric sensor // 1st Regional Symposium on Electrochemistry of South-East Europe / Horvat-Radošević, Višnja ; Mandić, Zoran ; Gojo, Miroslav (ur.).
Zagreb: Hrvatsko društvo kemijskih inženjera i tehnologa (HDKI), 2008. str. 288-289 (poster, domaća recenzija, sažetak, znanstveni)


CROSBI ID: 348002 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Determination of cationic surfactants in pharmaceutical disinfectants using a new, highly sensitive potentiometric sensor

Autori
Madunić-Čačić, Dubravka ; Sak-Bosnar, Milan ; Galović, Olivera ; Sakač, Nikola ; Matešić-Puač, Ružica ; Grabarić, Zorana

Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, znanstveni

Izvornik
1st Regional Symposium on Electrochemistry of South-East Europe / Horvat-Radošević, Višnja ; Mandić, Zoran ; Gojo, Miroslav - Zagreb : Hrvatsko društvo kemijskih inženjera i tehnologa (HDKI), 2008, 288-289

ISBN
978-953-6894-33-8

Skup
1st Regional Symposium on Electrochemistry of South-East Europe

Mjesto i datum
Rovinj, Hrvatska, 04.05.2008. - 08.05.2008

Vrsta sudjelovanja
Poster

Vrsta recenzije
Domaća recenzija

Ključne riječi
Potentiometric sensor; Cationic surfactants; Pharmaceutical disinfectants

Sažetak
A new sensitive potentiometric surfactant sensor was prepared based on a highly lipophilic 1, 3-didecyl-2-methyl-imidazolium (DMI) cation and a tetraphenylborate (TPB) antagonist ion. This compound was used as a sensing material and incorporated into the plasticized PVC-membrane. The sensor responded fast and showed a Nernstian response for all the investigated cationic surfactants: 59.8 mV/decade for cetylpyridinium chloride (CPC), 58.6 mV/decade for hexadecyltrimethylammonium bromide (CTAB), and 56.8 mV/decade for Hyamine. The sensor served as an end-point detector in ion-pair surfactant potentiometric titrations using sodium tetraphenylborate as titrant, and displayed satisfactory analytical performances within a pH range of two to eleven. The influence of the ethoxylated alcohols (EONS), a widely used class of nonionic surfactants on the shape of titration curves was negligible if the mass ratio of EONS:CS was not greater than five. The selectivity coefficients were determined by fitting the Nikolskii-Eisenman equation to the experimental data obtained by the mixed solution method. The sensor exhibited excellent selectivity performances for CPC over all organic and inorganic cations investigated. Several commercial disinfectant products containing various cationic surfactants were potentiometrically titrated using the new DMI-TPB sensor as an indicator.

Izvorni jezik
Engleski

Znanstvena područja
Kemija



POVEZANOST RADA


Projekti:
058-0580000-3071 - Razvoj kemometričkih i senzorskih metoda određivanja različitih analita (Ninčević Grassino, Antonela, MZOS ) ( CroRIS)
291-0580000-0169 - Kemijski senzori za primjenu u biomedicini, hrani i zaštiti okoliša (Sak-Bosnar, Milan, MZOS ) ( CroRIS)

Ustanove:
Prehrambeno-biotehnološki fakultet, Zagreb


Citiraj ovu publikaciju:

Madunić-Čačić, Dubravka; Sak-Bosnar, Milan; Galović, Olivera; Sakač, Nikola; Matešić-Puač, Ružica; Grabarić, Zorana
Determination of cationic surfactants in pharmaceutical disinfectants using a new, highly sensitive potentiometric sensor // 1st Regional Symposium on Electrochemistry of South-East Europe / Horvat-Radošević, Višnja ; Mandić, Zoran ; Gojo, Miroslav (ur.).
Zagreb: Hrvatsko društvo kemijskih inženjera i tehnologa (HDKI), 2008. str. 288-289 (poster, domaća recenzija, sažetak, znanstveni)
Madunić-Čačić, D., Sak-Bosnar, M., Galović, O., Sakač, N., Matešić-Puač, R. & Grabarić, Z. (2008) Determination of cationic surfactants in pharmaceutical disinfectants using a new, highly sensitive potentiometric sensor. U: Horvat-Radošević, V., Mandić, Z. & Gojo, M. (ur.)1st Regional Symposium on Electrochemistry of South-East Europe.
@article{article, author = {Maduni\'{c}-\v{C}a\v{c}i\'{c}, Dubravka and Sak-Bosnar, Milan and Galovi\'{c}, Olivera and Saka\v{c}, Nikola and Mate\v{s}i\'{c}-Pua\v{c}, Ru\v{z}ica and Grabari\'{c}, Zorana}, year = {2008}, pages = {288-289}, keywords = {Potentiometric sensor, Cationic surfactants, Pharmaceutical disinfectants}, isbn = {978-953-6894-33-8}, title = {Determination of cationic surfactants in pharmaceutical disinfectants using a new, highly sensitive potentiometric sensor}, keyword = {Potentiometric sensor, Cationic surfactants, Pharmaceutical disinfectants}, publisher = {Hrvatsko dru\v{s}tvo kemijskih in\v{z}enjera i tehnologa (HDKI)}, publisherplace = {Rovinj, Hrvatska} }
@article{article, author = {Maduni\'{c}-\v{C}a\v{c}i\'{c}, Dubravka and Sak-Bosnar, Milan and Galovi\'{c}, Olivera and Saka\v{c}, Nikola and Mate\v{s}i\'{c}-Pua\v{c}, Ru\v{z}ica and Grabari\'{c}, Zorana}, year = {2008}, pages = {288-289}, keywords = {Potentiometric sensor, Cationic surfactants, Pharmaceutical disinfectants}, isbn = {978-953-6894-33-8}, title = {Determination of cationic surfactants in pharmaceutical disinfectants using a new, highly sensitive potentiometric sensor}, keyword = {Potentiometric sensor, Cationic surfactants, Pharmaceutical disinfectants}, publisher = {Hrvatsko dru\v{s}tvo kemijskih in\v{z}enjera i tehnologa (HDKI)}, publisherplace = {Rovinj, Hrvatska} }




Contrast
Increase Font
Decrease Font
Dyslexic Font