Reliability of breath alcohol testing with Dräger Alcotest 7410Plus analyzer in a court process


Kovačić, Zdravko; Nestić, Marina; Stemberga, Valter; Bosnar, Alan; Petrovečki, Marko; Sutlović, Davorka
Reliability of breath alcohol testing with Dräger Alcotest 7410Plus analyzer in a court process // Medica Jadertina, 38 (2008), 1-2; 47-51 (podatak o recenziji nije dostupan, članak, ostalo)


CROSBI ID: 340220 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Reliability of breath alcohol testing with Dräger Alcotest 7410Plus analyzer in a court process

Autori
Kovačić, Zdravko ; Nestić, Marina ; Stemberga, Valter ; Bosnar, Alan ; Petrovečki, Marko ; Sutlović, Davorka

Izvornik
Medica Jadertina (0351-0093) 38 (2008), 1-2; 47-51

Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, ostalo

Ključne riječi
Dräger Alcotest 7410Plus; breath alcohol concentration (BrAC); blood alcohol concentration (BAC); forensic toxicology; court process

Sažetak
To determine usefulness and reliability of breath alcohol tests performed with Dräger Alcotest 7410Plus analyzer, in comparison to more complex analysis methods of blood and urine alcohol testing. A total of 570 cases were taken from archival court evidence material. In all cases individuals were driving under the influence of alcohol (DUI), and breath alcohol tests performed with Dräger Alcotest 7410Plus analyzer were positive. After breath alcohol testing, all individuals underwent repeated blood and urine alcohol testing, but this time concentrations were measured by headspace gas chromatography. In all individuals alcohol was in its elimination phase. Results obtained from both experiments were evaluated and compared. Comparison of breath alcohol concentrations and blood alcohol concentrations calculated to the time of breath alcotest, showed that 30 of 570 individuals (5.3%) were not within 95% predicament range, according to the linear equation CBAC ± 0.307 = 0.093 + 1.022CBrAC. Legal classification of beeing under the influence of alcohol refers to blood alcohol concentration, not breath alcohol concentration. Therefore it is not plausible to define the state of alcohol influence using Dräger Alcotest 7410Plus analyzer.

Izvorni jezik
Engleski

Znanstvena područja
Temeljne medicinske znanosti, Kliničke medicinske znanosti, Javno zdravstvo i zdravstvena zaštita



POVEZANOST RADA


Ustanove:
Medicinski fakultet, Rijeka,
Medicinski fakultet, Zagreb,
KBC Split,
Medicinski fakultet, Split

Poveznice na cjeloviti tekst rada:

Pristup cjelovitom tekstu rada

Citiraj ovu publikaciju:

Kovačić, Zdravko; Nestić, Marina; Stemberga, Valter; Bosnar, Alan; Petrovečki, Marko; Sutlović, Davorka
Reliability of breath alcohol testing with Dräger Alcotest 7410Plus analyzer in a court process // Medica Jadertina, 38 (2008), 1-2; 47-51 (podatak o recenziji nije dostupan, članak, ostalo)
Kovačić, Z., Nestić, M., Stemberga, V., Bosnar, A., Petrovečki, M. & Sutlović, D. (2008) Reliability of breath alcohol testing with Dräger Alcotest 7410Plus analyzer in a court process. Medica Jadertina, 38 (1-2), 47-51.
@article{article, author = {Kova\v{c}i\'{c}, Zdravko and Nesti\'{c}, Marina and Stemberga, Valter and Bosnar, Alan and Petrove\v{c}ki, Marko and Sutlovi\'{c}, Davorka}, year = {2008}, pages = {47-51}, keywords = {Dr\"{a}ger Alcotest 7410Plus, breath alcohol concentration (BrAC), blood alcohol concentration (BAC), forensic toxicology, court process}, journal = {Medica Jadertina}, volume = {38}, number = {1-2}, issn = {0351-0093}, title = {Reliability of breath alcohol testing with Dr\"{a}ger Alcotest 7410Plus analyzer in a court process}, keyword = {Dr\"{a}ger Alcotest 7410Plus, breath alcohol concentration (BrAC), blood alcohol concentration (BAC), forensic toxicology, court process} }
@article{article, author = {Kova\v{c}i\'{c}, Zdravko and Nesti\'{c}, Marina and Stemberga, Valter and Bosnar, Alan and Petrove\v{c}ki, Marko and Sutlovi\'{c}, Davorka}, year = {2008}, pages = {47-51}, keywords = {Dr\"{a}ger Alcotest 7410Plus, breath alcohol concentration (BrAC), blood alcohol concentration (BAC), forensic toxicology, court process}, journal = {Medica Jadertina}, volume = {38}, number = {1-2}, issn = {0351-0093}, title = {Reliability of breath alcohol testing with Dr\"{a}ger Alcotest 7410Plus analyzer in a court process}, keyword = {Dr\"{a}ger Alcotest 7410Plus, breath alcohol concentration (BrAC), blood alcohol concentration (BAC), forensic toxicology, court process} }

Časopis indeksira:


  • Scopus


Uključenost u ostale bibliografske baze podataka::


  • EMBASE (Excerpta Medica)





Contrast
Increase Font
Decrease Font
Dyslexic Font