Pretražite po imenu i prezimenu autora, mentora, urednika, prevoditelja

Napredna pretraga

Pregled bibliografske jedinice broj: 323880

A new high sensitive potentiometric sensor for anionic surfactants


Matešić-Puač, Ružica; Sak-Bosnar, Milan; Madunić-Čačić, Dubravka; Grabarić, Zorana
A new high sensitive potentiometric sensor for anionic surfactants // Proceedings of the 4th Croatian Symposium on Electrochemistry / Gojo, Miroslav ; Mandić, Zoran (ur.).
Primošten: Hrvatska, 2006. str. 221-227 (predavanje, međunarodna recenzija, cjeloviti rad (in extenso), znanstveni)


CROSBI ID: 323880 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
A new high sensitive potentiometric sensor for anionic surfactants

Autori
Matešić-Puač, Ružica ; Sak-Bosnar, Milan ; Madunić-Čačić, Dubravka ; Grabarić, Zorana

Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni

Izvornik
Proceedings of the 4th Croatian Symposium on Electrochemistry / Gojo, Miroslav ; Mandić, Zoran - Primošten : Hrvatska, 2006, 221-227

ISBN
953-6894-27-0

Skup
Hrvatski simpozij o elektrokemiji (4 ; 2006)

Mjesto i datum
Primošten, Hrvatska, 28.05.2006. - 01.06.2006

Vrsta sudjelovanja
Predavanje

Vrsta recenzije
Međunarodna recenzija

Ključne riječi
potentiometric sensor; anionic surfactants

Sažetak
A new anionic surfactant sensor has been prepared, based on 1, 3-didecyl-2-methyl-imidazolium-tetraphenylborate ion-exchange complex as sensing material incorporated into the plasticized PVC-membrane. The sensor exhibited Nernstian response (59.0 mV/decade) in the region between 2-10000 micromol/dm3 for dodecylsulfate and 61.6 mV/decade between 1-2000 micromol/dm3 for dodecylbenzen-sulfonate. The main application of the sensor described was indication of the end-point in ion-pair surfactant potentiometric titrations. The electrode showed satisfactory analytical performances within pH range 2.5 - 11. The potentiometric selectivity coefficients were determined by fixed interferent method. The sensor was also applied for potentiometric titration of some two-components mixtures.

Izvorni jezik
Engleski

Znanstvena područja
Kemija



POVEZANOST RADA


Projekti:
291-0580000-0169 - Kemijski senzori za primjenu u biomedicini, hrani i zaštiti okoliša (Sak-Bosnar, Milan, MZOS ) ( CroRIS)


Citiraj ovu publikaciju:

Matešić-Puač, Ružica; Sak-Bosnar, Milan; Madunić-Čačić, Dubravka; Grabarić, Zorana
A new high sensitive potentiometric sensor for anionic surfactants // Proceedings of the 4th Croatian Symposium on Electrochemistry / Gojo, Miroslav ; Mandić, Zoran (ur.).
Primošten: Hrvatska, 2006. str. 221-227 (predavanje, međunarodna recenzija, cjeloviti rad (in extenso), znanstveni)
Matešić-Puač, R., Sak-Bosnar, M., Madunić-Čačić, D. & Grabarić, Z. (2006) A new high sensitive potentiometric sensor for anionic surfactants. U: Gojo, M. & Mandić, Z. (ur.)Proceedings of the 4th Croatian Symposium on Electrochemistry.
@article{article, author = {Mate\v{s}i\'{c}-Pua\v{c}, Ru\v{z}ica and Sak-Bosnar, Milan and Maduni\'{c}-\v{C}a\v{c}i\'{c}, Dubravka and Grabari\'{c}, Zorana}, year = {2006}, pages = {221-227}, keywords = {potentiometric sensor, anionic surfactants}, isbn = {953-6894-27-0}, title = {A new high sensitive potentiometric sensor for anionic surfactants}, keyword = {potentiometric sensor, anionic surfactants}, publisher = {Hrvatska}, publisherplace = {Primo\v{s}ten, Hrvatska} }
@article{article, author = {Mate\v{s}i\'{c}-Pua\v{c}, Ru\v{z}ica and Sak-Bosnar, Milan and Maduni\'{c}-\v{C}a\v{c}i\'{c}, Dubravka and Grabari\'{c}, Zorana}, year = {2006}, pages = {221-227}, keywords = {potentiometric sensor, anionic surfactants}, isbn = {953-6894-27-0}, title = {A new high sensitive potentiometric sensor for anionic surfactants}, keyword = {potentiometric sensor, anionic surfactants}, publisher = {Hrvatska}, publisherplace = {Primo\v{s}ten, Hrvatska} }




Contrast
Increase Font
Decrease Font
Dyslexic Font