Pretražite po imenu i prezimenu autora, mentora, urednika, prevoditelja

Napredna pretraga

Pregled bibliografske jedinice broj: 321593

A potentnial high-throughput method for the determination of lipase activity by potentiometric flow injection titrations


Vahl, Katja; Kahlert, Heike; Bottcher, Dominique; Wardenga, Rainer; Komorsky-Lovrić, Šebojka; Scholz, Fritz
A potentnial high-throughput method for the determination of lipase activity by potentiometric flow injection titrations // Analytica chimica acta, 610 (2008), 1; 44-49 doi:10.1016/j.aca.2008.01.014 (međunarodna recenzija, članak, znanstveni)


CROSBI ID: 321593 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
A potentnial high-throughput method for the determination of lipase activity by potentiometric flow injection titrations

Autori
Vahl, Katja ; Kahlert, Heike ; Bottcher, Dominique ; Wardenga, Rainer ; Komorsky-Lovrić, Šebojka ; Scholz, Fritz

Izvornik
Analytica chimica acta (0003-2670) 610 (2008), 1; 44-49

Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni

Ključne riječi
flow injection titrations ; high-throughput screening ; lipase activity ; composite pH-detector
(titrations ; high-throughput screening ; lipase activity ; composite pH-detector)

Sažetak
Potentiometric FIA titrations were performed to determine enzyme activities of lipase type B from Candida antarctica, CAL.B. Two substrates, triacetin and tributyrin were hydrolyzed in phosphate buffer solutions, and the concentration change of the base component of the buffer was titrated in a carrier solution containing acid and potassium chloride. The system was calibated with butyric acid, respectivily. FIA titration peaks were evaluated with respect to pek high and peak area. Butyric acid and acetic acid could be titrated in the buffer solution from 3 x 10-3 molL-1 to 0.1 moljL-1. The detection limit enzyme activiti was determined to be 0.07 UmL-1 (15 min reaction time) and the minimum activity was calculated to be 0.035 units corresponding to 35 nmolmin-1. The specific acitivities of lipase B for the hydolysis of tributyrin and triacetin were determined as 16 +- 2 U mg-1 and 2 +- 0.2 U mg-1 (per mg commercial lipase preparation), respectively.

Izvorni jezik
Engleski

Znanstvena područja
Kemija



POVEZANOST RADA


Projekti:
098-0982904-2907 - Elektroanalitička istraživanja mikrokristala i tragova otopljenih supstanci (Lovrić, Milivoj, MZOS ) ( CroRIS)

Ustanove:
Institut "Ruđer Bošković", Zagreb

Profili:

Avatar Url Šebojka Komorsky-Lovrić (autor)

Poveznice na cjeloviti tekst rada:

doi www.sciencedirect.com doi.org

Citiraj ovu publikaciju:

Vahl, Katja; Kahlert, Heike; Bottcher, Dominique; Wardenga, Rainer; Komorsky-Lovrić, Šebojka; Scholz, Fritz
A potentnial high-throughput method for the determination of lipase activity by potentiometric flow injection titrations // Analytica chimica acta, 610 (2008), 1; 44-49 doi:10.1016/j.aca.2008.01.014 (međunarodna recenzija, članak, znanstveni)
Vahl, K., Kahlert, H., Bottcher, D., Wardenga, R., Komorsky-Lovrić, Š. & Scholz, F. (2008) A potentnial high-throughput method for the determination of lipase activity by potentiometric flow injection titrations. Analytica chimica acta, 610 (1), 44-49 doi:10.1016/j.aca.2008.01.014.
@article{article, author = {Vahl, Katja and Kahlert, Heike and Bottcher, Dominique and Wardenga, Rainer and Komorsky-Lovri\'{c}, \v{S}ebojka and Scholz, Fritz}, year = {2008}, pages = {44-49}, DOI = {10.1016/j.aca.2008.01.014}, keywords = {flow injection titrations, high-throughput screening, lipase activity, composite pH-detector}, journal = {Analytica chimica acta}, doi = {10.1016/j.aca.2008.01.014}, volume = {610}, number = {1}, issn = {0003-2670}, title = {A potentnial high-throughput method for the determination of lipase activity by potentiometric flow injection titrations}, keyword = {flow injection titrations, high-throughput screening, lipase activity, composite pH-detector} }
@article{article, author = {Vahl, Katja and Kahlert, Heike and Bottcher, Dominique and Wardenga, Rainer and Komorsky-Lovri\'{c}, \v{S}ebojka and Scholz, Fritz}, year = {2008}, pages = {44-49}, DOI = {10.1016/j.aca.2008.01.014}, keywords = {titrations, high-throughput screening, lipase activity, composite pH-detector}, journal = {Analytica chimica acta}, doi = {10.1016/j.aca.2008.01.014}, volume = {610}, number = {1}, issn = {0003-2670}, title = {A potentnial high-throughput method for the determination of lipase activity by potentiometric flow injection titrations}, keyword = {titrations, high-throughput screening, lipase activity, composite pH-detector} }

Časopis indeksira:


  • Current Contents Connect (CCC)
  • Web of Science Core Collection (WoSCC)
    • Science Citation Index Expanded (SCI-EXP)
    • SCI-EXP, SSCI i/ili A&HCI
  • Scopus
  • MEDLINE


Citati:





    Contrast
    Increase Font
    Decrease Font
    Dyslexic Font