Sažetak
The potential response of a symmetrical configuration in which the LaF3-membrance is placed between two solutions is discussed. The electrode body provides contact with the inner surface of the fluoride membrane, with a solution containing Fe(CN)63- - Fe(CN)64- redox couple and a Pt wire as internal reference electrode. The electrode was examined in terms of potential-concentration curves and potential-time response and shown to behave similarly to the commerical Orion fluoride electrode. The advantage of the proposed redox reference system is that the electrode has minimal drift immediately after assembly.