Pregled bibliografske jedinice broj: 27822
The effect of apical foramen and electrode diameter on the accuracy of electronic root canal measuring devices
The effect of apical foramen and electrode diameter on the accuracy of electronic root canal measuring devices // Proceedings of the 8th International IMEKO Conference on Measurement in Clinical Medicine, BMI'98 / Magjarević, Ratko (ur.).
Zagreb: Hrvatsko društvo za komunikacije, računarstvo, elektroniku, mjerenja I automatiku (KoREMA), 1998. (predavanje, međunarodna recenzija, cjeloviti rad (in extenso), znanstveni)
CROSBI ID: 27822 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
The effect of apical foramen and electrode diameter on the accuracy of electronic root canal measuring devices
Autori
Stare, Zoran ; Šutalo, Jozo ; Galić, Nada
Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni
Izvornik
Proceedings of the 8th International IMEKO Conference on Measurement in Clinical Medicine, BMI'98
/ Magjarević, Ratko - Zagreb : Hrvatsko društvo za komunikacije, računarstvo, elektroniku, mjerenja I automatiku (KoREMA), 1998
Skup
8th International IMEKO Conference on Measurement in Clinical Medicine, BMI'98
Mjesto i datum
Dubrovnik, Hrvatska, 16.09.1998. - 19.09.1998
Vrsta sudjelovanja
Predavanje
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
root canal length measurement; electronic apex locator; endodontics; bioelectric impedance
Sažetak
The influence of root canal diameter and file thickness on the magnitude of electric impedance, used to determine the length of the root canal by electronic apex locators, was tested on a model. As found, the magnitude of impedance depends on the diameter of the apical foramen and not on file size. The phenomena has been explained theoretically. It has also been demonstrated that the measurement actually involves the resistance of the electrolyte, which depends on the diameter of the apical foramen, because current density is the highest at that point. This also explains the different resistance change (sensitivity) for different file thickness and apical foramen diameters if the file tip is outside or inside the root canal.
Izvorni jezik
Engleski
Znanstvena područja
Elektrotehnika