Measurement of Electric Field Probe Error for Pulsed Signals
Measurement of Electric Field Probe Error for Pulsed Signals // 2006 IEEE International Symposium on Electromagnetic Compatibility
Portland (ME): Institute of Electrical and Electronics Engineers (IEEE), 2006. str. 244-248 (predavanje, međunarodna recenzija, cjeloviti rad (in extenso), znanstveni)
CROSBI ID: 249117 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Measurement of Electric Field Probe Error for Pulsed Signals
Autori
Šarolić, Antonio ; Roje, Vesna ; Modlic, Borivoj
Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni
Izvornik
2006 IEEE International Symposium on Electromagnetic Compatibility
/ - Portland (ME) : Institute of Electrical and Electronics Engineers (IEEE), 2006, 244-248
ISBN
1-4244-0293-X
Skup
2006 IEEE International Symposium on Electromagnetic Compatibility
Mjesto i datum
Portland (OR), Sjedinjene Američke Države, 14.08.2006. - 18.08.2006
Vrsta sudjelovanja
Predavanje
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
electric field probe; measurement error; pulsed signals; calibration waveguide
Sažetak
Electric field probe performance is measured inside a calibration waveguide. The electric field probes based on diode detection are designed and calibrated for continuous signals. When measuring electric field strength of a pulsed waveform, performance degradation is expected. The degradation is measured in terms of measurement error. It is defined as electric field strength reading on the probe vs. actual electric field strength in the test volume of a calibration waveguide. Two such probes were measured. Test signals were pulsed (ASK) signals with fixed duty cycle of 1/10 and varying repetition frequency. The results show great differences among the two probes measured. One probe shows only slight degradation in terms of increased reading (positive measurement error) for high repetition frequencies. The second probe shows the same feature but of even greater extent, together with significant degradation for low repetition frequencies in terms of decreased reading (negative measurement error). The existence of such errors demands a great amount of caution during EMC measurements using diode-based electric field probes.