Pregled bibliografske jedinice broj: 2079
Površinske doze pacijenata pri različitim CT pretragama
Površinske doze pacijenata pri različitim CT pretragama // Zbornik radova Trećeg simpozija Hrvatskog društva za zaštitu od zračenja / Kubelka, Dragan ; Kovač, Jadranka (ur.).
Zagreb: Hrvatsko društvo za zaštitu od zračenja, 1996. str. 237-243 (predavanje, nije recenziran, cjeloviti rad (in extenso), znanstveni)
CROSBI ID: 2079 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Površinske doze pacijenata pri različitim CT pretragama
(Patient surface doses in computerized tomography examinations)
Autori
Vekić, Branko ; Kovačević, Sandra ; Ranogajec- Komor, Maria ; Duvnjak, Neda ; Marušić, P. ; Anić, P. ; Dolenčić, P.
Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni
Izvornik
Zbornik radova Trećeg simpozija Hrvatskog društva za zaštitu od zračenja
/ Kubelka, Dragan ; Kovač, Jadranka - Zagreb : Hrvatsko društvo za zaštitu od zračenja, 1996, 237-243
Skup
Treći simpozij Hrvatskog društva za zaštitu od zračenja
Mjesto i datum
Zagreb, Hrvatska, 20.11.1996. - 22.11.1996
Vrsta sudjelovanja
Predavanje
Vrsta recenzije
Nije recenziran
Ključne riječi
površinske doze; pacijent; CT pretrage
(surface doses; patient; computerized tomography examinations)
Sažetak
Computed tomography (CT) has become a major source of the population exposure to diagnostic X-rays, and acknowledge of the doses delivered by the CT equipment has become very important. COnsiderable efforts should be made to keep these doses to a reasonable minimum, without sacrificing the image quality. The conditions of exposure in CT are quite different from dose in conventional X-ray imaging. This has required the development of specific techniques for assessing patient dose from CT. The aims of this work were to determine the dose delivered to various organs of patients undergoing computed tomography of abdomen, thorax, pelvis and kideny as measured on the surface of the body and to estimate the risk to the patients. Dosimetric measurements were performed at two different CT scanners (Siemens SOMATOM DR-H ve.r HC-1 and Shimadzu SCT-4500TE). The doses absorbed by different organs (gonads, chest, thyroid and eye lens) and by the examined part of the body of 95 patients of various sex and age were measured with TLD-700. The doses absorbed by different organs during the diagnostic CT examination of the body depend on the technical parameters, such as the number of scans, mAs, the thickness of scans, scanning times, tube voltage and other characteristics, some of each depend on the type and severity of illness. Clinical parameters, such as patient size and composition, and patient cooperation with regard to the control and motion, also influence the dose and the image quality. The highest dose measured in this study (89.19 mGy) was delivered to kidney during CT examination of this organ.
Izvorni jezik
Hrvatski
Znanstvena područja
Kemija
POVEZANOST RADA