Pregled bibliografske jedinice broj: 182806
Maintenance of the reference resistance standards of the Primary Electromagnetic Laboratory in Croatia
Maintenance of the reference resistance standards of the Primary Electromagnetic Laboratory in Croatia // Proceedings of the 13th IMEKO TC4 Symposium / Kayafas, E. ; Loumos, V. (ur.).
Atena: IMEKO&NTUA, 2004. str. 86-92 (predavanje, međunarodna recenzija, cjeloviti rad (in extenso), znanstveni)
CROSBI ID: 182806 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Maintenance of the reference resistance standards of the Primary Electromagnetic Laboratory in Croatia
Autori
Ilić, Damir ; Krešić, Mario ;
Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni
Izvornik
Proceedings of the 13th IMEKO TC4 Symposium
/ Kayafas, E. ; Loumos, V. - Atena : IMEKO&NTUA, 2004, 86-92
Skup
13th IMEKO TC4 Symposium
Mjesto i datum
Atena, Grčka, 29.09.2004. - 01.10.2004
Vrsta sudjelovanja
Predavanje
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
resistance standard ; least-squares method ; ultrathermostat
Sažetak
The past calibrations of the reference resistance standards of the Primary Electromagnetic Laboratory (PEL), the Leeds&Northrup (L&N) 1 ohm and 10 kohm standards, carried out over a period of more than thirty years, have been analysed. The least-squares fitting was used and the regression coefficients of up to the 3rd order polynomial were calculated and evaluated. Both standards have been maintaining into the self-developed oil ultrathermostat, where a stable temperature of 23 °C is maintained by the computer-based regulator within the limits of 10 mK. In the paper the analysis of the standards itself, as well as the achievement in the temperature regulation are described and pointed out.
Izvorni jezik
Engleski
Znanstvena područja
Elektrotehnika