Pregled bibliografske jedinice broj: 162696
Determination of Pu, Am and U in liquid samples by low-level alpha-spectrometry
Determination of Pu, Am and U in liquid samples by low-level alpha-spectrometry // XVIII Congress of Chemists and Technologists of Macedonia
Ohrid, 2004. (poster, nije recenziran, sažetak, znanstveni)
CROSBI ID: 162696 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Determination of Pu, Am and U in liquid samples by low-level alpha-spectrometry
Autori
Lulić, Stipe ; Mikelić, Luka ; Oreščanin, Višnja
Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, znanstveni
Izvornik
XVIII Congress of Chemists and Technologists of Macedonia
/ - Ohrid, 2004
Skup
XVIII Congress of Chemists and Technologists of Macedonia
Mjesto i datum
Ohrid, Sjeverna Makedonija, 23.09.2004. - 25.09.2004
Vrsta sudjelovanja
Poster
Vrsta recenzije
Nije recenziran
Ključne riječi
Pu; Am; U; Electrodeposition; Actinides; Alpha-spectrometry
Sažetak
A rapid, sensitive and selective procedure for determination of Pu, Am, and U in environmental and industrial liquid samples by electrodeposition technique was described. Optimum conditions for electrodeposition of actinides (plating time and current) were determined. Actinides were plated from ammonium oxalate-ammonium sulfate electrolyte containing diethyl triamino pentaacetic acid (DTPA). The optimized method was applied to environmental samples. The chemical yield of plated actinides was determined by 232U tracer. The maximum yield was obtained for electrodeposition lasted 3 hours and current 500 mA.
Izvorni jezik
Engleski
Znanstvena područja
Kemija