Pregled bibliografske jedinice broj: 158265
On-site monitoring of trace levels of free manganese in sea water via sono-electroanalysis using a boron-doped diamond electrode
On-site monitoring of trace levels of free manganese in sea water via sono-electroanalysis using a boron-doped diamond electrode // Analytica chimica acta, 533 (2005), 2; 141-145 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 158265 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
On-site monitoring of trace levels of free manganese in sea water via sono-electroanalysis using a boron-doped diamond electrode
Autori
Goodwin, Alexander ; Lawrence, Andrew ; Banks, Craig ; Wantz, Frédéric ; Omanović, Dario ; Komorsky-Lovrić, Šebojka ; Compton, Richard
Izvornik
Analytica chimica acta (0003-2670) 533
(2005), 2;
141-145
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
Manganese; Cathodic stripping voltammetry; Boron-doped diamond electrodes; Sea water; Environmental monitoring
Sažetak
Cathodic stripping voltammetry at a boron-doped diamond electrode utilising an insonated accumulation protocol is shown to be possible for the on-site trace detection of free manganese in seawater samples. No sample pre-treatment is required with the application of ultrasound providing a sensitive and selective protocol for the analysis of manganese. The protocol is used in the evaluation of manganese in seawater samples taken near the site of a former ferromanganese factory in the town Šibenik, Croatia.
Izvorni jezik
Engleski
Znanstvena područja
Geologija, Kemija
POVEZANOST RADA
Ustanove:
Institut "Ruđer Bošković", Zagreb
Citiraj ovu publikaciju:
Časopis indeksira:
- Current Contents Connect (CCC)
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus
Uključenost u ostale bibliografske baze podataka::
- Chemical Abstracts