TLD-700 for proton dosimetry in presence of low energy X-rays
TLD-700 for proton dosimetry in presence of low energy X-rays // Nuclear Instruments and Methods in Physics Research. Section A, 519 (2004), 667-673 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 144185 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
TLD-700 for proton dosimetry in presence of low energy X-rays
Autori
Miljanić, Saveta ; Ranogajec-Komor, Maria ; Blagus, Saša ; Miljanić, Đuro ; Osvay, M.
Izvornik
Nuclear Instruments and Methods in Physics Research. Section A (0168-9002) 519
(2004);
667-673
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
Proton dosimetry; Thermoluminescence dosimeters; TLD-700; CaF2:Mn; TL efficiency
Sažetak
The results of TLD-700 irradiations with 3 MeV protons using Tandem Van de Graaff accelerator are shown. In order to estimate the contribution of X-rays, CaF_2:Mn TL detectors were also irradiated at the same condition as TLD-700. Due to the presence of low energy X-rays, the peak area method of high-temperature peak 7 of TLD-700 was found to be more suitable for proton dose determination. TL efficiency for protons using this peak relatively to ^137Cs gamma rays was calculated. The relative responses (peak 7)_p/(peak 5)_gamma and (peak 7)_p/(peak 7)_gamma of 0.16 and 9.6, respectively were obtained. Taking into account the self-absorption of the TL photons in the dosimeter the relative efficiency for peak 7, ni_p, gamma, = 8.0 was found.
Izvorni jezik
Engleski
Znanstvena područja
Fizika, Kemija
POVEZANOST RADA
Ustanove:
Institut "Ruđer Bošković", Zagreb
Profili:
Đuro Miljanić
(autor)
Saveta Miljanić
(autor)
Ferencne Ranogajec
(autor)
Saša Blagus
(autor)
Citiraj ovu publikaciju:
Časopis indeksira:
- Current Contents Connect (CCC)
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus