Design of additive manufactured reference block with disk reflectors for sensitivity calibration of ultrasonic angle beam probes


Mihaljević, Morana; Markučič, Damir; Katić, Marko; Grozdanić, Maria; Kožić, Karlo
Design of additive manufactured reference block with disk reflectors for sensitivity calibration of ultrasonic angle beam probes // The International Journal of Advanced Manufacturing Technology, 126 (2023), 5-6; 1907-1918 doi:10.1007/s00170-023-11214-6 (međunarodna recenzija, članak, znanstveni)


CROSBI ID: 1278598 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Design of additive manufactured reference block with disk reflectors for sensitivity calibration of ultrasonic angle beam probes

Autori
Mihaljević, Morana ; Markučič, Damir ; Katić, Marko ; Grozdanić, Maria ; Kožić, Karlo

Izvornik
The International Journal of Advanced Manufacturing Technology (0268-3768) 126 (2023), 5-6; 1907-1918

Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni

Ključne riječi
Reference block ; Ultrasonic testing ; Sensitivity calibration ; Pulse-Echo ; DGS diagram ; Additive manufacturing

Sažetak
In this paper, a prototype of a reference block with disk reflectors for sensitivity calibration of ultrasonic testing system with angle beam probes was designed and produced. This paper aims to investigate the possibility for sensitivity calibration of ultrasonic angle beam probes with disk reflectors, with flat reference surfaces. The reference blocks with three artificially produced disk reflectors, located at angles of 45°, 60°, and 70°, were made by additive manufacturing technology direct metal laser sintering (DMLS), since DMLS has provided the possibility of producing components with complex inner geometries, which traditional processes could not produce. The exact size of the reference block is obtained by using CT scanning, and the equivalent reflector size (ERS) is estimated by the distance gain size (DGS) diagram and using DGS mode of the ultrasonic instrument by Pulse-Echo. The results are compared with each other and with both the nominal values and the exact size, obtained by CT scan.

Izvorni jezik
Engleski

Znanstvena područja
Strojarstvo



POVEZANOST RADA


Ustanove:
Fakultet strojarstva i brodogradnje, Zagreb

Poveznice na cjeloviti tekst rada:

doi link.springer.com www.webofscience.com

Citiraj ovu publikaciju:

Mihaljević, Morana; Markučič, Damir; Katić, Marko; Grozdanić, Maria; Kožić, Karlo
Design of additive manufactured reference block with disk reflectors for sensitivity calibration of ultrasonic angle beam probes // The International Journal of Advanced Manufacturing Technology, 126 (2023), 5-6; 1907-1918 doi:10.1007/s00170-023-11214-6 (međunarodna recenzija, članak, znanstveni)
Mihaljević, M., Markučič, D., Katić, M., Grozdanić, M. & Kožić, K. (2023) Design of additive manufactured reference block with disk reflectors for sensitivity calibration of ultrasonic angle beam probes. The International Journal of Advanced Manufacturing Technology, 126 (5-6), 1907-1918 doi:10.1007/s00170-023-11214-6.
@article{article, author = {Mihaljevi\'{c}, Morana and Marku\v{c}i\v{c}, Damir and Kati\'{c}, Marko and Grozdani\'{c}, Maria and Ko\v{z}i\'{c}, Karlo}, year = {2023}, pages = {1907-1918}, DOI = {10.1007/s00170-023-11214-6}, keywords = {Reference block, Ultrasonic testing, Sensitivity calibration, Pulse-Echo, DGS diagram, Additive manufacturing}, journal = {The International Journal of Advanced Manufacturing Technology}, doi = {10.1007/s00170-023-11214-6}, volume = {126}, number = {5-6}, issn = {0268-3768}, title = {Design of additive manufactured reference block with disk reflectors for sensitivity calibration of ultrasonic angle beam probes}, keyword = {Reference block, Ultrasonic testing, Sensitivity calibration, Pulse-Echo, DGS diagram, Additive manufacturing} }
@article{article, author = {Mihaljevi\'{c}, Morana and Marku\v{c}i\v{c}, Damir and Kati\'{c}, Marko and Grozdani\'{c}, Maria and Ko\v{z}i\'{c}, Karlo}, year = {2023}, pages = {1907-1918}, DOI = {10.1007/s00170-023-11214-6}, keywords = {Reference block, Ultrasonic testing, Sensitivity calibration, Pulse-Echo, DGS diagram, Additive manufacturing}, journal = {The International Journal of Advanced Manufacturing Technology}, doi = {10.1007/s00170-023-11214-6}, volume = {126}, number = {5-6}, issn = {0268-3768}, title = {Design of additive manufactured reference block with disk reflectors for sensitivity calibration of ultrasonic angle beam probes}, keyword = {Reference block, Ultrasonic testing, Sensitivity calibration, Pulse-Echo, DGS diagram, Additive manufacturing} }