Based on the constant cycle length (382 ms) of the re-VT during the prolongation of ventriculoatrial conduction (from 170 to 240 ms), the re-VT was most probably microreentrant. Different retrograde conduction times of the VT (160 ms) and PMT (315 ms) might suggest the existence of a dual retrograde conduction with separate pathways for each of the 2 tachycardias, but any evidence for such speculation is lacking. In any case, this case illustrates the complexity of possible interactions between spontaneous and device- mediated arrhythmias.