Pregled bibliografske jedinice broj: 122181
Second-generation Hybrid capture test and Amplicor monitor test generate highly correlated hepatitis B virus DNA levels
Second-generation Hybrid capture test and Amplicor monitor test generate highly correlated hepatitis B virus DNA levels // Journal of virological methods, 97 (2001), 1-2; 165-9 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 122181 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Second-generation Hybrid capture test and Amplicor monitor test generate highly correlated hepatitis B virus DNA levels
Autori
Poljak, M. ; Marin, I.J. ; Seme, K. ; Brinovec, V. ; Maticic, M. ; Meglic-Volkar, J. ; Lesnicar, G. ; Vince, A.
Izvornik
Journal of virological methods (0166-0934) 97
(2001), 1-2;
165-9
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
hepatitis B
(hepatitsi B)
Sažetak
The performance of the Digene Hybrid Capture II HBV DNA Test HC II and the Roche Cobas Amplicor Monitor Test (Cobas-HBV) was evaluated on 252 serum samples. One hundred and seventy-three samples were HBV DNA positive and 75 HBV DNA negative by both assays. Four samples were HBV DNA positive by Cobas-HBV only. Linear regression analysis showed that the HBV DNA concentrations obtained from both assays were significantly related (n=173, r=0.976, P<0.0001). The results of the study show that Hybrid capture II and Cobas-HBV could be used equally in the management for patients with chronic HBV infection.
Izvorni jezik
Engleski
Znanstvena područja
Kliničke medicinske znanosti
Citiraj ovu publikaciju:
Časopis indeksira:
- Current Contents Connect (CCC)
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus
- MEDLINE
Uključenost u ostale bibliografske baze podataka::
- Index Medicus