ANALYTICAL VERIFICATION AND QUALITY ASSESSMENT OF A FULLY AUTOMATED IMMUNOTURBIDIMETRIC HBA1C DETERMINATION USING HBA1C ADVANCED KIT ON BECKMAN COULTER DXC 700 AU ANALYTICAL SYSTEM


Božičević, Sandra; Penava, Filip; Vučić Lovrenčić, Marijana; Krhač Grgić, Maja; Matišić, Ena; Kardum Paro Mirjana Mariana; Smirčić Duvnjak, Lea
ANALYTICAL VERIFICATION AND QUALITY ASSESSMENT OF A FULLY AUTOMATED IMMUNOTURBIDIMETRIC HBA1C DETERMINATION USING HBA1C ADVANCED KIT ON BECKMAN COULTER DXC 700 AU ANALYTICAL SYSTEM // "Other" Clinical Chemistry and Laboratory Medicine (CCLM), vol. 59, no. s1, 2021, pp. s820-s864.
München, Njemačka: Walter de Gruyter, 2021. str. 94-998 doi:10.1515/cclm-2021-5032 (poster, međunarodna recenzija, sažetak, znanstveni)


CROSBI ID: 1191522 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
ANALYTICAL VERIFICATION AND QUALITY ASSESSMENT OF A FULLY AUTOMATED IMMUNOTURBIDIMETRIC HBA1C DETERMINATION USING HBA1C ADVANCED KIT ON BECKMAN COULTER DXC 700 AU ANALYTICAL SYSTEM

Autori
Božičević, Sandra ; Penava, Filip ; Vučić Lovrenčić, Marijana ; Krhač Grgić, Maja ; Matišić, Ena ; Kardum Paro Mirjana Mariana ; Smirčić Duvnjak, Lea

Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, znanstveni

Izvornik
"Other" Clinical Chemistry and Laboratory Medicine (CCLM), vol. 59, no. s1, 2021, pp. s820-s864. / - : Walter de Gruyter, 2021, 94-998

Skup
24th IFCC-EFLM European Congress of Clinical Chemistry and Laboratory Medicine (EuroMedlab 2021)

Mjesto i datum
München, Njemačka, 10.04.2022. - 14.04.2022

Vrsta sudjelovanja
Poster

Vrsta recenzije
Međunarodna recenzija

Ključne riječi
verification HbA1c ; DXC 700

Sažetak
BACKGROUND-AIM. The clinical indications for HbA1c measurement represent a challenge in meeting rigorous quality requirements. The aim of the study was analytical verification of HbA1c Advanced reagent (BCA) on DxC 700 AU analyser (both by BeckmanCoulter, USA) for HbA1c measurement in whole blood samples and to asses clinical reliability of results in comparison to two accredited methods. METHODS. Total imprecision was verified according to CLSI EP15-A2 protocol using commercial quality control (CQC) (ExtendSURE, Canterbury Scientific, NZ) and pooled whole blood (WB) samples. Bias was estimated with External quality control samples (ERL, 2021.1-19 ; values determined by the reference method in European Reference Laboratory for Glycohemoglobin Standardization, The Netherlands). Results were compared to Tina-quant, Gen 3 immunoassay (IM) (Cobas Integra 400+, Roche, USA) (N=174) and capillaryelectrophoresis (CE)(Minicap Flex- Piercing, Sebia, France)(N=28) on patient samples remaining after routine HbA1c analysis. RESULTS. Total imprecision of BCA method for CQC was 2, 64% (HbA1c 37 mmol/mol) and 2, 02% (HbA1c 86 mmol/mol) ; and for WB 1, 07% (HbA1c 59 mmol/mol). The deviation of the results from target reference values range was (-3 to 1 mmol/mol)[average deviation: -0.67 mmol/mol absolute, 1.42% relative]. Passing-Bablok regression showed excellent method correlation with neither constant nor proportional difference (regression equation: y=0, 9000+0, 9714x ; intercept A=0, 9000, 95% CI=-1, 0000-1, 8182 ; slope B=0, 9714 ; 95% CI=0, 9545- 1, 0000 ; Spearman's corelation coefficient 0, 997). A minimal, clinically insignificant constant difference between BCA and CE was found, with excellent correlation in study on 28 samples of pregnant women with gestational diabetes (regression eqation: Y=-1, 0000+1, 0000x ; intercept A=-1, 0000 ; 95% CI=-6, 4167 to -1, 0000 ; slope B=1, 0000 ; 95% CI=1, 0000-1, 1667). Sigma metrics result σ=4 indicates BCA meet strict quality requirements, at recommended clinical decision level 5 mmol/mol (IFCC Task Force for HbA1c). Verification of reference intervals confirmed the compliance with recommended WHO values with HbA1c intervals of 16- 36 mmol/mol. CONCLUSIONS. The HbA1c Advanced Kit method on Beckman Coulter DxC 700 AU meets all rigorous quality criteria for reliable clinical use in research, diagnosis and monitoring of diabetes

Izvorni jezik
Engleski

Znanstvena područja
Biotehnologija u biomedicini (prirodno područje, biomedicina i zdravstvo, biotehničko područje)



POVEZANOST RADA


Ustanove:
Medicinski fakultet, Zagreb

Poveznice na cjeloviti tekst rada:

doi www.degruyter.com

Citiraj ovu publikaciju:

Božičević, Sandra; Penava, Filip; Vučić Lovrenčić, Marijana; Krhač Grgić, Maja; Matišić, Ena; Kardum Paro Mirjana Mariana; Smirčić Duvnjak, Lea
ANALYTICAL VERIFICATION AND QUALITY ASSESSMENT OF A FULLY AUTOMATED IMMUNOTURBIDIMETRIC HBA1C DETERMINATION USING HBA1C ADVANCED KIT ON BECKMAN COULTER DXC 700 AU ANALYTICAL SYSTEM // "Other" Clinical Chemistry and Laboratory Medicine (CCLM), vol. 59, no. s1, 2021, pp. s820-s864.
München, Njemačka: Walter de Gruyter, 2021. str. 94-998 doi:10.1515/cclm-2021-5032 (poster, međunarodna recenzija, sažetak, znanstveni)
Božičević, S., Penava, F., Vučić Lovrenčić, M., Krhač Grgić, M., Matišić, E., Kardum Paro Mirjana Mariana & Smirčić Duvnjak, L. (2021) ANALYTICAL VERIFICATION AND QUALITY ASSESSMENT OF A FULLY AUTOMATED IMMUNOTURBIDIMETRIC HBA1C DETERMINATION USING HBA1C ADVANCED KIT ON BECKMAN COULTER DXC 700 AU ANALYTICAL SYSTEM. U: "Other" Clinical Chemistry and Laboratory Medicine (CCLM), vol. 59, no. s1, 2021, pp. s820-s864. doi:10.1515/cclm-2021-5032.
@article{article, author = {Bo\v{z}i\v{c}evi\'{c}, Sandra and Penava, Filip and Vu\v{c}i\'{c} Lovren\v{c}i\'{c}, Marijana and Krha\v{c} Grgi\'{c}, Maja and Mati\v{s}i\'{c}, Ena and Smir\v{c}i\'{c} Duvnjak, Lea}, year = {2021}, pages = {94-998}, DOI = {10.1515/cclm-2021-5032}, keywords = {verification HbA1c, DXC 700}, doi = {10.1515/cclm-2021-5032}, title = {ANALYTICAL VERIFICATION AND QUALITY ASSESSMENT OF A FULLY AUTOMATED IMMUNOTURBIDIMETRIC HBA1C DETERMINATION USING HBA1C ADVANCED KIT ON BECKMAN COULTER DXC 700 AU ANALYTICAL SYSTEM}, keyword = {verification HbA1c, DXC 700}, publisher = {Walter de Gruyter}, publisherplace = {M\"{u}nchen, Njema\v{c}ka} }
@article{article, author = {Bo\v{z}i\v{c}evi\'{c}, Sandra and Penava, Filip and Vu\v{c}i\'{c} Lovren\v{c}i\'{c}, Marijana and Krha\v{c} Grgi\'{c}, Maja and Mati\v{s}i\'{c}, Ena and Smir\v{c}i\'{c} Duvnjak, Lea}, year = {2021}, pages = {94-998}, DOI = {10.1515/cclm-2021-5032}, keywords = {verification HbA1c, DXC 700}, doi = {10.1515/cclm-2021-5032}, title = {ANALYTICAL VERIFICATION AND QUALITY ASSESSMENT OF A FULLY AUTOMATED IMMUNOTURBIDIMETRIC HBA1C DETERMINATION USING HBA1C ADVANCED KIT ON BECKMAN COULTER DXC 700 AU ANALYTICAL SYSTEM}, keyword = {verification HbA1c, DXC 700}, publisher = {Walter de Gruyter}, publisherplace = {M\"{u}nchen, Njema\v{c}ka} }

Časopis indeksira:


  • Current Contents Connect (CCC)
  • Web of Science Core Collection (WoSCC)
    • Science Citation Index Expanded (SCI-EXP)
    • SCI-EXP, SSCI i/ili A&HCI
  • Scopus
  • MEDLINE


Citati:





    Contrast
    Increase Font
    Decrease Font
    Dyslexic Font