A Hygrometer Calibration Facility Based on Two Temperature Humidity Generator


Zvizdić, Davor; Galaso, Ivan; Šerfezi, Damir; Kolobarić, Vladimir
A Hygrometer Calibration Facility Based on Two Temperature Humidity Generator // Papers and Abstracts from the Third International Symposium on Humidity&Moisture - Volume 1 / Stephanie, Bell (ur.).
Teddington: NPL - National Physical Laboratory, 1998. str. 119-126 (poster, međunarodna recenzija, cjeloviti rad (in extenso), znanstveni)


CROSBI ID: 11878 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
A Hygrometer Calibration Facility Based on Two Temperature Humidity Generator

Autori
Zvizdić, Davor ; Galaso, Ivan ; Šerfezi, Damir ; Kolobarić, Vladimir

Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni

Izvornik
Papers and Abstracts from the Third International Symposium on Humidity&Moisture - Volume 1 / Stephanie, Bell - Teddington : NPL - National Physical Laboratory, 1998, 119-126

Skup
Third International Symposium on Humidity&Moisture

Mjesto i datum
Teddington, Ujedinjeno Kraljevstvo, 06.04.1998. - 08.04.1998

Vrsta sudjelovanja
Poster

Vrsta recenzije
Međunarodna recenzija

Ključne riječi
higrometri;kalibracija
(hygrometers; calibration)

Sažetak
Abstract: To respond to the increasing demand for traceable calibration of hygrometers in Croatia, a hygrometer calibration facility has been developed. It is based on two temperature recirculating humidity generator intended for operation in the 4°C to 60°C dew point range. The design of the facility is a compromise one, having to accomodate dew-point hygrometers, relative humidity meters and even chart recorders without remote sensors, i.e. the need for different calibration routines and levels of accuracy. The facility can , therefore, operate in three conceptualy different modes: as stand alone humidity generator, as a slave system directly controlled by transfer standard hygrometer and as an environmental chamber for comparison calibrations. The whole system is integrated and automated via the custom developed, graphycally interfaced software platform. The main design features of the facility are described as well as considerations regarding behaviour and the accuracy of the system.

Izvorni jezik
Engleski

Znanstvena područja
Strojarstvo



POVEZANOST RADA


Projekti:
120007
120043

Ustanove:
Fakultet strojarstva i brodogradnje, Zagreb

Profili:

Avatar Url Damir Šerfezi (autor)

Avatar Url Ivan Galaso (autor)

Avatar Url Vladimir Kolobarić (autor)


Citiraj ovu publikaciju:

Zvizdić, Davor; Galaso, Ivan; Šerfezi, Damir; Kolobarić, Vladimir
A Hygrometer Calibration Facility Based on Two Temperature Humidity Generator // Papers and Abstracts from the Third International Symposium on Humidity&Moisture - Volume 1 / Stephanie, Bell (ur.).
Teddington: NPL - National Physical Laboratory, 1998. str. 119-126 (poster, međunarodna recenzija, cjeloviti rad (in extenso), znanstveni)
Zvizdić, D., Galaso, I., Šerfezi, D. & Kolobarić, V. (1998) A Hygrometer Calibration Facility Based on Two Temperature Humidity Generator. U: Stephanie, B. (ur.)Papers and Abstracts from the Third International Symposium on Humidity&Moisture - Volume 1.
@article{article, author = {Zvizdi\'{c}, Davor and Galaso, Ivan and \v{S}erfezi, Damir and Kolobari\'{c}, Vladimir}, editor = {Stephanie, B.}, year = {1998}, pages = {119-126}, keywords = {higrometri, kalibracija}, title = {A Hygrometer Calibration Facility Based on Two Temperature Humidity Generator}, keyword = {higrometri, kalibracija}, publisher = {NPL - National Physical Laboratory}, publisherplace = {Teddington, Ujedinjeno Kraljevstvo} }
@article{article, author = {Zvizdi\'{c}, Davor and Galaso, Ivan and \v{S}erfezi, Damir and Kolobari\'{c}, Vladimir}, editor = {Stephanie, B.}, year = {1998}, pages = {119-126}, keywords = {hygrometers, calibration}, title = {A Hygrometer Calibration Facility Based on Two Temperature Humidity Generator}, keyword = {hygrometers, calibration}, publisher = {NPL - National Physical Laboratory}, publisherplace = {Teddington, Ujedinjeno Kraljevstvo} }