Pregled bibliografske jedinice broj: 1145711
Clinical application of the RONNA G4 system – preliminary validation of 23 robotic frameless brain biopsies
Clinical application of the RONNA G4 system – preliminary validation of 23 robotic frameless brain biopsies // Croatian medical journal, 62 (2021), 4; 318-327 doi:10.3325/cmj.2021.62.318 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 1145711 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Clinical application of the RONNA G4 system – preliminary validation of 23
robotic frameless brain biopsies
Autori
Dlaka, Domagoj ; Švaco, Marko ; Chudy, Darko ; Jerbić, Bojan ; Šekoranja, Bojan ; Šuligoj, Filip ; Vidaković, Josip ; Almahariq, Fadi ; Romić, Dominik ; Raguž, Marina
Izvornik
Croatian medical journal (0353-9504) 62
(2021), 4;
318-327
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
RONNA G4 ; neuronavigation ; robotics ; neurosurgery
Sažetak
Aims: To report the outcomes of robot-assisted brain biopsies performed using a novel RONNA G4 system. The system was developed by a research group from the Faculty of Mechanical Engineering and Naval Architecture and a team of neurosurgeons from Dubrava University Hospital, University of Zagreb School of Medicine. Methods: This prospective study included 49 biopsies analyzed during one year: 23 robotic frameless and 26 frame-based Leksell stereotactic biopsies. We analyzed the presenting symptoms, tumor range and location, postoperative complications, pathohistological diagnosis, diagnostic yield, as well as operation and hospitalization duration. The target point error was calculated to assess the accuracy of the RONNA system. Results: No postoperative mortality, morbidity, or infections were observed. In the frameless robotic biopsy group, only one pathohistological diagnosis was inconclusive. Therefore, the diagnostic yield was 95.6% (22/23), similar to that of the framebased Leksell stereotactic biopsy group (95.1% or 25/26). The average target point error in the frameless robotic biopsy group was 2.15±1.22 mm (range 0.39-5.85). Conclusion: The RONNA G4 robotic system is a safe and accurate tool for brain biopsy, although further research warrants a larger patient sample, comparison with other robotic systems, and a systematic analysis of the entry and target point errors.
Izvorni jezik
Engleski
Znanstvena područja
Temeljne tehničke znanosti, Kliničke medicinske znanosti
POVEZANOST RADA
Ustanove:
Medicinski fakultet, Zagreb,
Fakultet strojarstva i brodogradnje, Zagreb,
Klinička bolnica "Dubrava"
Profili:
Bojan Jerbić
(autor)
Josip Vidaković
(autor)
Marko Švaco
(autor)
Filip Šuligoj
(autor)
Darko Chudy
(autor)
Bojan Šekoranja
(autor)
Fadi Almahariq
(autor)
Marina Raguž
(autor)
Citiraj ovu publikaciju:
Časopis indeksira:
- Current Contents Connect (CCC)
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus
- MEDLINE