Time-dependent variation of ionized calcium in serum samples


Perovic, Antonija; Njire Braticevic, Marina
Time-dependent variation of ionized calcium in serum samples // Biochemia medica, 29 (2019), 3; 1-9 doi:10.11613/BM.2019.030708 (međunarodna recenzija, članak, stručni)


CROSBI ID: 1096645 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Time-dependent variation of ionized calcium in serum samples

Autori
Perovic, Antonija ; Njire Braticevic, Marina

Izvornik
Biochemia medica (1330-0962) 29 (2019), 3; 1-9

Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, stručni

Ključne riječi
ionized calcium ; serum ; heparin ; specimen handling ; pre-analytical phase

Sažetak
Introduction: The aim of this study was to compare ionized calcium (iCa) concentrations in arterial heparinized blood and venous serum and to investigate time-dependent variation of iCa in serum samples centrifuged and analysed at different times. Materials and methods: Ionized calcium was measured (N = 25) in arterial blood within 20 min after puncture, and in serum within 10 min after centrifugation conducted 30 min after sampling. Effect of time between sampling and centrifugation was examined in three tubes (N = 30) centrifuged 15, 30 and 60 min after sampling, and analysed within 10 min. Effect of time between centrifugation and analysis was investigated in three tubes (N = 31) centrifuged 30 min after sampling and analysed: 0-10, 30-40 and 90-100 min after centrifugation. Ionized calcium was measured on the Siemens RapidLab 348EX analysers. Statistical significance was tested using Wilcoxon test and ANOVA analysis. Clinical significance was judged against reference change values (RCV). Results: No statistically significant difference was found between iCa in arterial blood and serum (P = 0.274). A statistically significant decrease was found: in tubes centrifuged 60 and 15 min after sampling versus 30 min (P = 0.005, P = 0.003) ; and in tubes analysed 30-40 and 90-100 min after centrifugation versus 0-10 min (P = 0.021, P = 0.027). Clinically significant changes were observed: 60 versus 30 min (centrifugation) and 90-100 versus 0-10 and 30-40 min (analysis). Conclusions: Timely analysed arterial blood and serum samples can be used interchangeably. To avoid clinically significant variations, serum tubes should be centrifuged within 30 min after sampling, and analysis should be performed within 30 min after centrifugation.

Izvorni jezik
Engleski



POVEZANOST RADA


Poveznice na cjeloviti tekst rada:

doi

Citiraj ovu publikaciju:

Perovic, Antonija; Njire Braticevic, Marina
Time-dependent variation of ionized calcium in serum samples // Biochemia medica, 29 (2019), 3; 1-9 doi:10.11613/BM.2019.030708 (međunarodna recenzija, članak, stručni)
Perovic, A. & Njire Braticevic, M. (2019) Time-dependent variation of ionized calcium in serum samples. Biochemia medica, 29 (3), 1-9 doi:10.11613/BM.2019.030708.
@article{article, author = {Perovic, Antonija and Njire Braticevic, Marina}, year = {2019}, pages = {1-9}, DOI = {10.11613/BM.2019.030708}, keywords = {ionized calcium, serum, heparin, specimen handling, pre-analytical phase}, journal = {Biochemia medica}, doi = {10.11613/BM.2019.030708}, volume = {29}, number = {3}, issn = {1330-0962}, title = {Time-dependent variation of ionized calcium in serum samples}, keyword = {ionized calcium, serum, heparin, specimen handling, pre-analytical phase} }
@article{article, author = {Perovic, Antonija and Njire Braticevic, Marina}, year = {2019}, pages = {1-9}, DOI = {10.11613/BM.2019.030708}, keywords = {ionized calcium, serum, heparin, specimen handling, pre-analytical phase}, journal = {Biochemia medica}, doi = {10.11613/BM.2019.030708}, volume = {29}, number = {3}, issn = {1330-0962}, title = {Time-dependent variation of ionized calcium in serum samples}, keyword = {ionized calcium, serum, heparin, specimen handling, pre-analytical phase} }

Časopis indeksira:


  • Current Contents Connect (CCC)
  • Web of Science Core Collection (WoSCC)
    • Science Citation Index Expanded (SCI-EXP)
    • SCI-EXP, SSCI i/ili A&HCI
  • Scopus
  • MEDLINE


Citati:





    Contrast
    Increase Font
    Decrease Font
    Dyslexic Font