Pregled bibliografske jedinice broj: 1075715
Final report on key comparison EURAMET.M.P-K1.c in the range 0.7 MPa to 7.0 MPa of gas gauge pressure
Final report on key comparison EURAMET.M.P-K1.c in the range 0.7 MPa to 7.0 MPa of gas gauge pressure // Metrologia, Volume 57 (2020), Number 1A 07022, 43 doi:10.1088/0026-1394/57/1A/07022 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 1075715 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Final report on key comparison EURAMET.M.P-K1.c
in
the range 0.7 MPa to 7.0 MPa of gas gauge
pressure
Autori
Aykurt, Altıntaş ; Ilknur, Koçaş ; Yasin, Durgut ; Joseph, Bartolo ; Mercede, Bergoglio ; Lovorka, Grgec Bermanec ; Alen, Bošnjaković ; Sanja, Burzić ; Antoine, Condereys ; Miruna, Dobre ; Peter, Farar ; Paul, Hetherington ; Nieves, Medina ; Wladimir, Sabuga ; Oliver, Ott ; Thomas, Konczak ; Ion, Sandu ; Janez, Setina ; Dietmar, Steindl ; Csilla, Vámossy ; Bernard, Waller ; Christian, Wuethrich ; Adam, Brzozowski ; Jan, van Geel ; Sari, Saxhol ; Fredrik, Arrhén ;
Izvornik
Metrologia (0026-1394) Volume 57
(2020);
Number 1A 07022, 43
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
gas pressure, intercomparison
Sažetak
A EURAMET key comparison of the national pressure standards in the range 0.7 MPa to 7.0 MPa of gas gauge pressure was carried out. The circulation of the transfer standard began in November 2011 and lasted until November 2016. The measurand of the comparison was the effective area of the piston-cylinder assembly determined by gauge pressure measurements in the range from 0.7 MPa to 7.0 MPa. As the comparison reference value, the weighted mean of the results of the laboratories with primary pressure standards was used. With this reference value, all the participants who delivered the results demonstrated equivalence respective to the reference value within expanded uncertainties (k = 2) on all the range. The results of this comparison were linked to CCM key comparison CCM.P-K1.c. Also in relation to the reference values of CCM.P- K1.c, all participants demonstrated agreement within expanded uncertainties (k = 2) at all pressure points.
Izvorni jezik
Engleski
Znanstvena područja
Fizika, Strojarstvo, Interdisciplinarne tehničke znanosti
POVEZANOST RADA
Ustanove:
Fakultet strojarstva i brodogradnje, Zagreb
Citiraj ovu publikaciju:
Časopis indeksira:
- Current Contents Connect (CCC)
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus