Haemolysis and lipemia interfere with resistin and myeloperoxidase BioVendor ELISA assays
Haemolysis and lipemia interfere with resistin and myeloperoxidase BioVendor ELISA assays // Biochemia Medica, 29 (2019), 2; 1-7 doi:10.11613/bm.2019.020703 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 1000470 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Haemolysis and lipemia interfere with resistin and myeloperoxidase BioVendor ELISA assays
Autori
Saracevic, Andrea ; Dukic, Lora ; Simundic, Ana- Maria
Izvornik
Biochemia Medica (1330-0962) 29
(2019), 2;
1-7
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
haemolysis ; lipemia ; interferences ; resistin ; myeloperoxidase
Sažetak
Introduction The aim of our study was to investigate the influence of haemolysis and lipemia on resistin (RES) and myeloperoxidase (MPO) measurement by BioVendor enzyme-linked immunosorbent assays (ELISA). Materials and methods Blood was taken from healthy volunteers into lithium heparin tubes. Plasma samples were spiked with Lipofundin® emulsion (B. Braun Melsungen AG, Germany) for lipemia interference testing. Haemolysed samples were obtained by drawing aliquots of heparinized blood through a 26 gauge needle. Index of haemolysis (H), lipemia (L) and triglyceride concentration were measured on Abbott Architect c8000. Haemoglobin concentration was measured on Sysmex XN-1000. Concentrations of RES and MPO in all samples were determined with RES and MPO ELISA kits (BioVendor, Czech Republic). All measurements were performed in triplicate. Biases from the native samples were calculated for both analytes and compared with an arbitrary value (e.g. ± 10%). Results Triglyceride concentration in the investigated samples ranged from 0.57 to 38.23 mmol/L, which corresponds to L index from - 0.01 to 13.77. Haemoglobin concentration in all samples ranged from 0 to 8 g/L which correspond to H index from 0.05 to 8.77. Both MPO and RES showed significant biases at 1 g/L haemoglobin (58.7% and 66.7%, respectively). Also, both MPO and RES showed significant biases at 4.66 mmol/L triglycerides (33.8% and - 12.2%, respectively). Conclusions Resistin BioVendor assays are affected by haemolysis and lipemia already at low degree of interferent. Haemolysis was found to interfere at 1 g/L haemoglobin for both assays, while lipemia interferes at 4.66 mmol/L of triglycerides.
Izvorni jezik
Engleski
Znanstvena područja
Temeljne medicinske znanosti, Kliničke medicinske znanosti
POVEZANOST RADA
Ustanove:
Farmaceutsko-biokemijski fakultet, Zagreb,
Klinička bolnica "Sveti Duh"
Citiraj ovu publikaciju:
Časopis indeksira:
- Current Contents Connect (CCC)
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus
- MEDLINE