Capillary blood sampling for POCT analysis: left vs. right hand
Capillary blood sampling for POCT analysis: left vs. right hand // Clinical Chemistry and Laboratory Medicine / Plebani, Mario (ur.).
Berlin: Walter de Gruyter, 2019. str. eA62-eA62 doi:10.1515/cclm-2019-0104 (poster, međunarodna recenzija, sažetak, stručni)
CROSBI ID: 990217 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Capillary blood sampling for POCT analysis: left vs. right hand
Autori
Božović, Marija ; Milevoj Kopčinović, Lara ; Culej, Jelena ; Vrtarić, Alen ; Topić, Anita ; Miler, Marijana ; Nikolac Gabaj, Nora
Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, stručni
Izvornik
Clinical Chemistry and Laboratory Medicine
/ Plebani, Mario - Berlin : Walter de Gruyter, 2019, EA62-eA62
Skup
5th EFLM Conference on Preanalytical Phase
Mjesto i datum
Zagreb, Hrvatska, 22.03.2019. - 23.03.2019
Vrsta sudjelovanja
Poster
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
capillary blood sampling, POCT, left hand, right hand
Sažetak
BACKGROUND-AIM: Point-of-care testing (POCT) comprises diagnostic testing usually performed by non-laboratory personnel at the site of the patient (i.e. hospital wards or at home). The comparability of results obtained from left hand (L) vs. right hand (R) capillary sampling samples using POCT is implied. Our aim was to test this hypothesis, i.e. to test if results obtained from L and R hand capillary samples by using a POCT device were comparable. METHODS: Paired capillary samples were taken simultaneously from L and R middle fingers of 21 healthy participants using Greiner bio-one miniCollect safety lancets with puncture depth of 2.00 mm. A volume of 200 μL of capillary blood from each finger was collected using lithium- heparinized Minivette POCT (Sarstedt, Nümbrecht, Germany). Testing was performed using Abbott i-STAT Alinity (Abbott, Abbott Park, USA) POCT analyser and Na, K, Cl, iCa, glucose, urea, creatinine, haematocrit, haemoglobin, pH, PCO2, PO2, TCO2, HCO3-, base excess (BE), sO2, lactate were measured. Statistical analysis was performed using MedCalc statistical software (Ostend, Belgium). Mean biases obtained from L vs. R were compared to CROQALM criteria for respective parameters. RESULTS: No statistical or clinical difference in results obtained after sampling from L and R was found. Calculated biases for each parameter analysed were lower than the criteria previously defined. The biases expressed as mean bias (corresponding criteria) were as follows: pH 0.04% (1%), pCO2 - 0.24% (8%), pO2 2.00% (8%), lactate -4.35% (16%), Na 0.30% (3%), K 1.65% (6%), Cl – 0.15% (4%), urea 0.88% (8%), creatinine 1.19% (9%), glucose - 0.87% (7%), iCa 1.79% (4%), Hct 1.25% (5%), Hb 1.22 (5%). CONCLUSIONS: Results obtained from L and R hand capillary samples using a POCT device were comparable. Thus, capillary sampling might be performed from either hand, interchangeably.
Izvorni jezik
Engleski
Znanstvena područja
Interdisciplinarne prirodne znanosti, Kliničke medicinske znanosti
POVEZANOST RADA
Ustanove:
KBC "Sestre Milosrdnice"
Profili:
Alen Vrtarić
(autor)
Anita Radman
(autor)
Nora Nikolac
(autor)
Marijana Miler
(autor)
Lara Milevoj Kopčinović
(autor)
Citiraj ovu publikaciju:
Časopis indeksira:
- Current Contents Connect (CCC)
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- Scopus
- MEDLINE