Pregled bibliografske jedinice broj: 984249
Internal quality assurance of HIL indices on Roche Cobas c702
Internal quality assurance of HIL indices on Roche Cobas c702 // PLoS One, 13 (2018), 7; 0200088, 11 doi:10.1371/journal.pone.0200088 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 984249 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Internal quality assurance of HIL indices on Roche Cobas c702
Autori
Lippi, Giuseppe ; Cadamuro, Janne ; Danese, Elisa ; Gelati, Matteo ; Montagnana, Martina ; von Meyer, Alexander ; Salvagno, Gian Luca ; Šimundić, Ana-Maria
Izvornik
PLoS One (1932-6203) 13
(2018), 7;
0200088, 11
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
HIL
Sažetak
Automatic assessment of hemoglobin (H), lipaemia (L) and icterus (I) in serum or plasma (HIL indices) is the mainstay for evaluating sample quality. We planned this study to verify whether in-house prepared internal quality control (IQC) materials may be suitable for quality assurance of HIL indices. Pools containing different values of each of the three HIL indices were prepared from routine plasma samples, divided in aliquots and frozen at -20°C. Stability of frozen materials was assessed by thawing one aliquot of each pool after different days of freezing (1, 4, 8, 15, 22 and 29), and by measuring HIL indices on baseline fresh samples and frozen-thawed aliquots with Roche Cobas c702. Five fresh liquid IQCs materials were also measured at the same time points. Intra-assay and inter-assay imprecision of HIL indices calculated with commercial IQC materials ranged between 1.1–2.0% and 1.6–3.3%, respectively. When target values of HIL indices were calculated using frozen-thawed aliquots, the inter-assay imprecision of in-house prepared materials was optimal, even better than that of commercial liquid IQCs (H-index, 0.8% versus 1.6% ; L-index, 2.2% versus 2.5% ; I-index, 0.8% versus 3.3%). In conclusion, in-house prepared IQC materials are cost-effective alternatives to commercial liquid IQCs for HIL quality assurance.
Izvorni jezik
Engleski
Znanstvena područja
Kliničke medicinske znanosti
Citiraj ovu publikaciju:
Časopis indeksira:
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus
- MEDLINE
Uključenost u ostale bibliografske baze podataka::
- AGRICOLA
- EMBASE (Excerpta Medica)
- FSTA: Food Science and Technology Abstracts
- GeoRef
- PsychINFO
- Zoological Record
- Google Scholar
- CAS
- PubMed