Dose profiles measurements for leksell Gamma- knife model C using automatic positioning system with stereotactic diode detector


Hršak, Hrvoje; Majer, Marija; Heinrich, Zdravko
Dose profiles measurements for leksell Gamma- knife model C using automatic positioning system with stereotactic diode detector // 8th AAMP conference Proceedings / Petrović, Borislava ; Ciraj-Bjelac, Olivera (ur.).
Novi Sad: Serbian Association of Medical Physicists, 2017. str. 149-156 (predavanje, međunarodna recenzija, cjeloviti rad (in extenso), znanstveni)


CROSBI ID: 899186 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Dose profiles measurements for leksell Gamma- knife model C using automatic positioning system with stereotactic diode detector

Autori
Hršak, Hrvoje ; Majer, Marija ; Heinrich, Zdravko

Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni

Izvornik
8th AAMP conference Proceedings / Petrović, Borislava ; Ciraj-Bjelac, Olivera - Novi Sad : Serbian Association of Medical Physicists, 2017, 149-156

ISBN
978-86-7306-145-0

Skup
8th Alpe Adria Medical Physics Meeting

Mjesto i datum
Novi Sad, Srbija, 25.05.2017. - 27.05.2017

Vrsta sudjelovanja
Predavanje

Vrsta recenzije
Međunarodna recenzija

Ključne riječi
Gamma-Knife radiosurgery ; Dose profiles ; Quality assurance ; Diode detector ; Radiochromic film

Sažetak
Gamma-Knife radiosurgery is highly precise method for treatment of small intracranial lesions with high single radiation dose [1]. For Leksell Gamma-Knife unit (LGK) Model C (Elekta AB, Stockholm, Sweden) radiosurgery uses 201 convergent narrow Co-60 photon beams, collimated with the 18, 14, 8 and 4 mm helmet collimator. The quality assurance program (QA) for Gamma- Knife includes periodical checks of size and position of single collimator helmet dose distributions from which complex clinical distributions are composed [2]. This is done by measuring dose distributions profiles in a three orthogonal planes (usually along x, y and z - axis of Gamma-Knife unit) and comparing them against the profiles measured during commissioning of Gamma-Knife or calculated by the treatment planning system (TPS). A method was proposed for high resolution dose profiles measurements for LGK Model C using the Automatic positioning system (APS) - a motorized system for positioning of the patient head and stereotactic frame into the target coordinates [7], with p-type stereotactic diode detector. Measurements were performed in a non- reference geometry where spherical phantom with detector was moved from one measuring point to another using APS. The method was validated by comparison of profiles simulated in a non- reference geometry of phantom against the profiles simulated in a reference geometry using the Leksell Gamma Plan (LGP) TPS and by comparison of profiles measured with EBT3 film with phantom in reference geometry against the profiles measured with the Diode SRS in a non- reference geometry. Diode SRS PTW60018 detector represent good choice for the Gamma-Knife dosimetry because of high spatial resolution and good signal response and in conjunction with the Automatic positioning system it can provide high resolution and volume averaging free dose profile measurement for the Leksell Gamma-Knife Model C. In addition, less noise in a measured signal and simpler workflow represent one advantage over the radiochromic film. However, EBT3 radiochromic film remains the detector of choice for the 2D dosimetry because of high spatial resolution and tissue equivalence.

Izvorni jezik
Engleski

Znanstvena područja
Fizika, Kliničke medicinske znanosti



POVEZANOST RADA


Ustanove:
Institut "Ruđer Bošković", Zagreb,
Klinički bolnički centar Zagreb

Profili:

Avatar Url Marija Majer (autor)

Avatar Url Hrvoje Hršak (autor)

Avatar Url Zdravko Heinrich (autor)

Poveznice na cjeloviti tekst rada:

Pristup cjelovitom tekstu rada

Citiraj ovu publikaciju:

Hršak, Hrvoje; Majer, Marija; Heinrich, Zdravko
Dose profiles measurements for leksell Gamma- knife model C using automatic positioning system with stereotactic diode detector // 8th AAMP conference Proceedings / Petrović, Borislava ; Ciraj-Bjelac, Olivera (ur.).
Novi Sad: Serbian Association of Medical Physicists, 2017. str. 149-156 (predavanje, međunarodna recenzija, cjeloviti rad (in extenso), znanstveni)
Hršak, H., Majer, M. & Heinrich, Z. (2017) Dose profiles measurements for leksell Gamma- knife model C using automatic positioning system with stereotactic diode detector. U: Petrović, B. & Ciraj-Bjelac, O. (ur.)8th AAMP conference Proceedings.
@article{article, author = {Hr\v{s}ak, Hrvoje and Majer, Marija and Heinrich, Zdravko}, year = {2017}, pages = {149-156}, keywords = {Gamma-Knife radiosurgery, Dose profiles, Quality assurance, Diode detector, Radiochromic film}, isbn = {978-86-7306-145-0}, title = {Dose profiles measurements for leksell Gamma- knife model C using automatic positioning system with stereotactic diode detector}, keyword = {Gamma-Knife radiosurgery, Dose profiles, Quality assurance, Diode detector, Radiochromic film}, publisher = {Serbian Association of Medical Physicists}, publisherplace = {Novi Sad, Srbija} }
@article{article, author = {Hr\v{s}ak, Hrvoje and Majer, Marija and Heinrich, Zdravko}, year = {2017}, pages = {149-156}, keywords = {Gamma-Knife radiosurgery, Dose profiles, Quality assurance, Diode detector, Radiochromic film}, isbn = {978-86-7306-145-0}, title = {Dose profiles measurements for leksell Gamma- knife model C using automatic positioning system with stereotactic diode detector}, keyword = {Gamma-Knife radiosurgery, Dose profiles, Quality assurance, Diode detector, Radiochromic film}, publisher = {Serbian Association of Medical Physicists}, publisherplace = {Novi Sad, Srbija} }




Contrast
Increase Font
Decrease Font
Dyslexic Font