Investigation of implantable signal transmission characteristics based on visible data of the human leg
Investigation of implantable signal transmission characteristics based on visible data of the human leg // Biomedical engineering online, 16 (2017), 88; 1-14 doi:10.1186/s12938-017-0379-y (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 884292 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Investigation of implantable signal transmission characteristics based on visible data of the human leg
Autori
Gao, Yue-Ming ; Ye, Yan-Ting ; Lin, Shi ; Lučev Vasić, Željka ; Vai, Mang-I ; Du, Min ; Cifrek, Mario ; Pun, Sio-Hang
Izvornik
Biomedical engineering online (1475-925X) 16
(2017), 88;
1-14
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
Channel modeling ; Finite element model ; Implantable signal transmission ; Phantom model
Sažetak
Background: Signal transmission characteristics between implanted medical devices and external equipment has been a common key issue, as has the problem of supplying energy to the devices. It can be used to enable signal transmission from implanted devices that the human body’s conductive properties. Using signal transmission by galvanic coupling is one of the most effective signal transmission methods. Methods: The signal transmission characteristics by galvanic coupling of implantable devices using a frequency range of 10 kHz to 1 MHz was analyzed in this article. A finite element (FEM) model and a phantom model established by visible human leg data were used to investigate the signal transmission characteristics of implant-to- surface, with implantable receiver electrodes at different locations. Results: The results showed that the FEM model and the phantom model had similar implantable signal transmission characteristics, with an increase of frequency, signal attenuation basically remained unchanged. The gain in signal attenuation in the fixed attenuation values fluctuated no more than 5 dB and signal attenuation values rose as the channel length increased. Conclusions: Our results of signal transmission characteristics of surface-to-implant will provide a theoretical basis for implantable transceiver design, and for realization of a recharging method for implanted medical devices.
Izvorni jezik
Engleski
Znanstvena područja
Elektrotehnika
POVEZANOST RADA
Ustanove:
Fakultet elektrotehnike i računarstva, Zagreb
Citiraj ovu publikaciju:
Časopis indeksira:
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus
- MEDLINE
Uključenost u ostale bibliografske baze podataka::
- EMBASE (Excerpta Medica)
- INSPEC
- Citebase
- DOAJ
- EmBiology
- EmCare
- OAIster
- SCImago
- SOCOLAR
- Zetoc