Cross-reactivity of serum indices on Abbott Architect c8000, Roche Cobas 6000 and Beckman Coulter AU5800 analyzers: HIL interference project of the Working group for the preanalytical phase, Croatian Society of Medical Biochemistry and Laboratory Medicine


Miler, Marijana; Nikolac, Nora; Ćelap, Ivana; Filipi, Petra; Hemar, Marina; Kocijančić, Marija; Šimundić, Ana-Maria; Šupak Smolčić, Vesna; Vrtarić, Alen
Cross-reactivity of serum indices on Abbott Architect c8000, Roche Cobas 6000 and Beckman Coulter AU5800 analyzers: HIL interference project of the Working group for the preanalytical phase, Croatian Society of Medical Biochemistry and Laboratory Medicine // Clinical Chemistry and Laboratory Medicine / Plebani, Mario (ur.).
Berlin: Walter de Gruyter, 2017. str. eA20-eA20 (poster, međunarodna recenzija, sažetak, znanstveni)


CROSBI ID: 883044 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Cross-reactivity of serum indices on Abbott Architect c8000, Roche Cobas 6000 and Beckman Coulter AU5800 analyzers: HIL interference project of the Working group for the preanalytical phase, Croatian Society of Medical Biochemistry and Laboratory Medicine

Autori
Miler, Marijana ; Nikolac, Nora ; Ćelap, Ivana ; Filipi, Petra ; Hemar, Marina ; Kocijančić, Marija ; Šimundić, Ana-Maria ; Šupak Smolčić, Vesna ; Vrtarić, Alen

Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, znanstveni

Izvornik
Clinical Chemistry and Laboratory Medicine / Plebani, Mario - Berlin : Walter de Gruyter, 2017, EA20-eA20

Skup
4th EFLM-BD European Conference on Preanalytical Phase

Mjesto i datum
Amsterdam, Nizozemska, 24.03.2017. - 25.03.2017

Vrsta sudjelovanja
Poster

Vrsta recenzije
Međunarodna recenzija

Ključne riječi
cross-reactivity, hemolysis, icteria, interference, lipemia, serum indices

Sažetak
Serum indices, hemolysis (H), icteria (I), lipemia (L) are determined automatically on chemistry analyzers. HIL absorbance spectra patterns overlap which could result with HIL cross-reactivity when multiple interferences are present. We aimed to investigate interference cross-reactivity and to compare cross-reactivity of indices between three different biochemistry analyzers. Study included Abbott Architect c8000, Beckman Coulter AU5800 (BC) and Roche Cobas 6000. For each interference, 6 increasing levels of interfering substances were added to serum pool (hemolysis: 0.3-10 g/L hemoglobin (Hb) by needle aspiration, icteria: 35-675 µmol/L of dissolved bilirubin, lipemia: 0.1-1.4 g/L of Intralipid). HIL indices were measured in duplicate before and after addition of second interferent (approximate values H=5 g/L, I=400 µmol/L, L=1 g/L of Intralipid) and biases between those measurements were calculated. BC provides semiquantitative results (5 categories), while Abbott and Roche express indices as corresponding concentrations: hemoglobin (g/L) for H, bilirubin (μmol/L) for I and Intralipid (g/L) for L index. Less than 10% was considered acceptable bias for Abbott and Roche, and for BC agreement of result category. Hemolysis decreased I index on Abbott and Roche (for 20% and 14%, respectively). On BC, hemolysis increased I index. L index on Abbott was higher (for 35-180%), while on Roche and BC, L index was lower (11-32% for Roche) after adding free Hb. Adding bilirubin had no effect on H index on Abbott, while on Roche and BC, H index was higher (13-35%). Bilirubin also changed L index for more than 30% on all analyzers. I index on Abbott and Roche was lower after adding Intralipid (mean bias 75% and 28%, respectively). Lipemia interfered with H index on all three analyzers (for approximately 14%). Almost all tested interference combinations showed cross- reactivity. Serum indices could be erroneous if more than one interfering substances is present in serum sample.

Izvorni jezik
Engleski

Znanstvena područja
Kliničke medicinske znanosti



POVEZANOST RADA


Ustanove:
KBC "Sestre Milosrdnice"

Citiraj ovu publikaciju:

Miler, Marijana; Nikolac, Nora; Ćelap, Ivana; Filipi, Petra; Hemar, Marina; Kocijančić, Marija; Šimundić, Ana-Maria; Šupak Smolčić, Vesna; Vrtarić, Alen
Cross-reactivity of serum indices on Abbott Architect c8000, Roche Cobas 6000 and Beckman Coulter AU5800 analyzers: HIL interference project of the Working group for the preanalytical phase, Croatian Society of Medical Biochemistry and Laboratory Medicine // Clinical Chemistry and Laboratory Medicine / Plebani, Mario (ur.).
Berlin: Walter de Gruyter, 2017. str. eA20-eA20 (poster, međunarodna recenzija, sažetak, znanstveni)
Miler, M., Nikolac, N., Ćelap, I., Filipi, P., Hemar, M., Kocijančić, M., Šimundić, A., Šupak Smolčić, V. & Vrtarić, A. (2017) Cross-reactivity of serum indices on Abbott Architect c8000, Roche Cobas 6000 and Beckman Coulter AU5800 analyzers: HIL interference project of the Working group for the preanalytical phase, Croatian Society of Medical Biochemistry and Laboratory Medicine. U: Plebani, M. (ur.)Clinical Chemistry and Laboratory Medicine.
@article{article, author = {Miler, Marijana and Nikolac, Nora and \'{C}elap, Ivana and Filipi, Petra and Hemar, Marina and Kocijan\v{c}i\'{c}, Marija and \v{S}imundi\'{c}, Ana-Maria and \v{S}upak Smol\v{c}i\'{c}, Vesna and Vrtari\'{c}, Alen}, editor = {Plebani, M.}, year = {2017}, pages = {eA20-eA20}, keywords = {cross-reactivity, hemolysis, icteria, interference, lipemia, serum indices}, title = {Cross-reactivity of serum indices on Abbott Architect c8000, Roche Cobas 6000 and Beckman Coulter AU5800 analyzers: HIL interference project of the Working group for the preanalytical phase, Croatian Society of Medical Biochemistry and Laboratory Medicine}, keyword = {cross-reactivity, hemolysis, icteria, interference, lipemia, serum indices}, publisher = {Walter de Gruyter}, publisherplace = {Amsterdam, Nizozemska} }
@article{article, author = {Miler, Marijana and Nikolac, Nora and \'{C}elap, Ivana and Filipi, Petra and Hemar, Marina and Kocijan\v{c}i\'{c}, Marija and \v{S}imundi\'{c}, Ana-Maria and \v{S}upak Smol\v{c}i\'{c}, Vesna and Vrtari\'{c}, Alen}, editor = {Plebani, M.}, year = {2017}, pages = {eA20-eA20}, keywords = {cross-reactivity, hemolysis, icteria, interference, lipemia, serum indices}, title = {Cross-reactivity of serum indices on Abbott Architect c8000, Roche Cobas 6000 and Beckman Coulter AU5800 analyzers: HIL interference project of the Working group for the preanalytical phase, Croatian Society of Medical Biochemistry and Laboratory Medicine}, keyword = {cross-reactivity, hemolysis, icteria, interference, lipemia, serum indices}, publisher = {Walter de Gruyter}, publisherplace = {Amsterdam, Nizozemska} }

Časopis indeksira:


  • Current Contents Connect (CCC)
  • Web of Science Core Collection (WoSCC)
    • Science Citation Index Expanded (SCI-EXP)
    • SCI-EXP, SSCI i/ili A&HCI
  • Scopus
  • MEDLINE





Contrast
Increase Font
Decrease Font
Dyslexic Font