Pregled bibliografske jedinice broj: 866509
Odstranjivanje frakturiranih vijaka implanta pomoću ultrazvučne sonde - prikaz slučaja
Odstranjivanje frakturiranih vijaka implanta pomoću ultrazvučne sonde - prikaz slučaja // Sažeci Međunarodnog kongresa Hrvatskog stomatološkog društva HLZ-a i 1. Međunarodnog kongresa Hrvatskog društva za minimalno intervencijsku dentalnu medicinu HLZ-a u suorganizaciji Stomatološkog fakulteta Sveučilišta u Zagrebu: „Znanost i umjetnost zajedno“. U: Acta Stomatologica Croatica 2017 ; 51(1) / Vidović Juras, Danica ; Jurić, Hrvoje ; Baraba, Anja (ur.).
Zagreb: Stomatolški fakultet Sveučilišta u Zagrebu, 2017. str. 83-83 (poster, međunarodna recenzija, sažetak, stručni)
CROSBI ID: 866509 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Odstranjivanje frakturiranih vijaka implanta pomoću
ultrazvučne sonde - prikaz slučaja
(Retrieval of fractured implant screws using
ultrasonic scaler - a case report)
Autori
Azizi, Bleron ; Mehmeti, Blerim, Matulić, Nena ; Gabrić, Dragana
Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, stručni
Izvornik
Sažeci Međunarodnog kongresa Hrvatskog stomatološkog društva HLZ-a i 1. Međunarodnog kongresa Hrvatskog društva za minimalno intervencijsku dentalnu medicinu HLZ-a u suorganizaciji Stomatološkog fakulteta Sveučilišta u Zagrebu: „Znanost i umjetnost zajedno“. U: Acta Stomatologica Croatica 2017 ; 51(1)
/ Vidović Juras, Danica ; Jurić, Hrvoje ; Baraba, Anja - Zagreb : Stomatolški fakultet Sveučilišta u Zagrebu, 2017, 83-83
Skup
Međunarodni kongres Hrvatskog stomatološkog društva HLZ-a i 1. Međunarodni kongres Hrvatskog društva za minimalno intervencijsku dentalnu medicinu HLZ-a
Mjesto i datum
Zagreb, Hrvatska, 18.11.2017. - 19.11.2017
Vrsta sudjelovanja
Poster
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
oral surgery ; dental implantology ; explantation
Sažetak
This case report presents a non-invasive method of removing two broken implant screws using ultrasonic scaler. A 49 years old male patient presented with a 3 unit screw retained bridge (14- 16) which had fallen out, due to screw fracture on both implants. Examination and intraoral x-rays showed two dental implants, in good position, with slight crestal bone resorption. The screws were fractured below the level of the hexagonal lock. Small incision was made in order to access the screws. A small pit was formed with a carbide bur. The ultrasonic scaler tip (Great Star G3, GGEE Co., Ltd, China) was placed in the pit and slow anti clockwise movement was applied, which was gradually increased. The implant screw at position 14 was retrieved after 2 minutes of anti-clockwise movement, and the implant screw at position 16 was retrieved after 4 minutes of movement. After removing the screws, a control x-ray was taken in order to confirm the removal of the screws. The internal surface of the implants was uncontaminated and the same screw retained bridge was placed with two new screws which were tightened up to 25 N/cm. No clinical problems were observed at 6 months follow- up.The use of Ultra sonic scaler can be a successful non-invasive method for retrieving broken implant screws. It is of great importance to avoid damage to the internal hex of the implants which can lead to prosthodontically unrestorable implants.
Izvorni jezik
Hrvatski, engleski
Znanstvena područja
Dentalna medicina
POVEZANOST RADA
Projekti:
MZOS-065-0000000-0424 - Cijeljenje koštanog defekta nakon imedijatne implantacije (Katanec, Davor, MZOS ) ( CroRIS)
Ustanove:
Stomatološki fakultet, Zagreb
Profili: